Comparison of naive sixth-grade children with trained professionals in the use of an automated external defibrillator

被引:158
作者
Gundry, JW [1 ]
Comess, KA [1 ]
DeRook, FA [1 ]
Jorgenson, D [1 ]
Bardy, GH [1 ]
机构
[1] Univ Washington, Med Ctr, Dept Med, Div Cardiol, Seattle, WA 98195 USA
关键词
defibrillation; fibrillation; death; sudden; cardiopulmonary resuscitation;
D O I
10.1161/01.CIR.100.16.1703
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background-Survival after out-of-hospital cardiac arrest (OHCA) is strongly influenced by time to defibrillation, Wider availability of automated external defibrillators (AEDs) may decrease response times but only with increased lay use. Consequently, this study endeavored to improve our understanding of AED use in naive users by measuring times to shock and appropriateness of pad location. We chose sixth-grade students to simulate an extreme circumstance of unfamiliarity with the problem of OHCA and defibrillation. The children's AED use was then compared with that of professionals. Methods and Results-With the use of a mock cardiac arrest scenario, AED use by 15 children was compared with that of 22 emergency medical technicians (EMTs) or paramedics. The primary end point was time from entry onto the cardiac arrest scene to delivery of the shock into simulated ventricular fibrillation, The secondary end point was appropriateness of pad placement. All subject performances were videotaped to assess safety of use and compliance with AED prompts to remain clear of the mannequin during shock delivery. Mean time to defibrillation was 90+/-14 seconds (range, 69 to 111 seconds) for the children and 67+/-10 seconds (range, 50 to 87 seconds) for the EMTs/paramedics (P<0.0001). Electrode pad placement was appropriate for all subjects. All remained clear of the "patient" during shock delivery, Conclusions-During mock cardiac arrest, the speed of AED use by untrained children is only modestly slower than that of professionals. The difference between the groups is surprisingly small, considering the naivete of the children as untutored first-time users. These findings suggest that widespread use of AEDs will require only modest training.
引用
收藏
页码:1703 / 1707
页数:5
相关论文
共 25 条
[1]  
*AM HEART ASS, 1996, EXP ACC DEF LEG ADV
[2]   Low-energy biphasic waveform defibrillation: Evidence-based review applied to emergency cardiovascular care guidelines - A statement for healthcare professionals from the American Heart Association Committee on Emergency Cardiovascular Care and the subcommittees on basic life support, advanced cardiac life support, and pediatric resuscitation [J].
Cummins, RO ;
Hazinski, MF ;
Kerber, RE ;
Kudenchuk, P ;
Becker, L ;
Nichol, G ;
Malanga, B ;
Aufderheide, TP ;
Stapleton, EM ;
Kern, K ;
Ornato, JP ;
Sanders, A ;
Valenzuela, T ;
Eisenberg, M .
CIRCULATION, 1998, 97 (16) :1654-1667
[3]   ENCOURAGING EARLY DEFIBRILLATION - THE AMERICAN-HEART-ASSOCIATION AND AUTOMATED EXTERNAL DEFIBRILLATORS [J].
CUMMINS, RO ;
THIES, W .
ANNALS OF EMERGENCY MEDICINE, 1990, 19 (11) :1245-1248
[4]   TRAINING LAY PERSONS TO USE AUTOMATIC EXTERNAL DEFIBRILLATORS - SUCCESS OF INITIAL TRAINING AND ONE-YEAR RETENTION OF SKILLS [J].
CUMMINS, RO ;
SCHUBACH, JA ;
LITWIN, PE ;
HEARNE, TR .
AMERICAN JOURNAL OF EMERGENCY MEDICINE, 1989, 7 (02) :143-149
[5]  
CUMMINS RO, 1997, CIRCULATION S1, V96, P365
[6]  
DIACK AW, 1979, MED INSTRUM, V13, P78
[7]   CARDIAC-ARREST AND RESUSCITATION - A TALE OF 29 CITIES [J].
EISENBERG, MS ;
HORWOOD, BT ;
CUMMINS, RO ;
REYNOLDSHAERTLE, R ;
HEARNE, TR .
ANNALS OF EMERGENCY MEDICINE, 1990, 19 (02) :179-186
[8]  
EWY GA, 1982, NEW ENGL J MED, V306, P174
[9]   SUDDEN CORONARY DEATH IN THE UNITED-STATES - 1980-1985 [J].
GILLUM, RF .
CIRCULATION, 1989, 79 (04) :756-765
[10]  
Gliner Bradford E., 1998, Biomedical Instrumentation and Technology, V32, P631