A comparison of power colour flow with frequency based colour flow Doppler in fetal echocardiography

被引:5
作者
Chua, LKL [1 ]
Twining, P [1 ]
机构
[1] QUEENS MED CTR,DEPT RADIOL,NOTTINGHAM NG7 2UH,ENGLAND
关键词
D O I
10.1016/S0009-9260(97)80038-4
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
In this study a direct comparison was made between power colour flow and frequency based colour flow Doppler in the visualization of the ventricular septum and the pulmonary veins of the fetal heart at 18 to 20 weeks, A total of 16 cases were scanned using both power colour flow and frequency based colour flow Doppler, The entire ventricular septum was visualized in all cases using power colour, but in only two cases using frequency based colour Doppler, However, frequency based colour Doppler was capable of identifying part of the ventricular septum in 14 cases, In the assessment of the pulmonary veins, power colour demonstrated both right and left pulmonary veins dearly in 12 cases, There were no cases in which both the right and left pulmonary veins were demonstrated using frequency based colour Doppler.
引用
收藏
页码:712 / 714
页数:3
相关论文
共 5 条
[1]  
ALLAN LD, 1984, BRIT HEART J, V52, P542
[2]   POWER VERSUS CONVENTIONAL COLOR DOPPLER SONOGRAPHY - COMPARISON IN THE DEPICTION OF NORMAL INTRARENAL VASCULATURE [J].
BUDE, RO ;
RUBIN, JM ;
ADLER, RS .
RADIOLOGY, 1994, 192 (03) :777-780
[3]  
COPEL JA, 1991, OBSTET GYNECOL, V78, P1
[4]   POWER DOPPLER US - A POTENTIALLY USEFUL ALTERNATIVE TO MEAN FREQUENCY-BASED COLOR DOPPLER US [J].
RUBIN, JM ;
BUDE, RO ;
CARSON, PL ;
BREE, RL ;
ADLER, RS .
RADIOLOGY, 1994, 190 (03) :853-856
[5]  
TWINING P, 1994, BMUS B, V2, P22