Comments on high-accuracy viscosity measurements using capillary viscometers

被引:6
作者
Einfeldt, J [1 ]
机构
[1] Univ Rostock, Dept Phys, D-18051 Rostock, Germany
关键词
D O I
10.1088/0026-1394/38/5/8
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
The working equation of a capillary viscometer, used generally in the form v = kv - B/tau (where v is kinematic viscosity; tau is outflow time; k and B are viscometer constants), can be deduced directly from the Navier-Stokes equation. The surface tension only affects the k parameter. For a Ubbelohde-type viscometer with suspended level, the curvature of the meniscus, which compensates the influence of the mean surface tension, has been calculated. The relation often used for the capillary pressure in tubes, p(sigma) = 2sigma/R, is only approximately valid for radii R < 0.3 mm. The Knibbs method, which uses external variable pressure, provides a means of calibrating viscometers to high accuracy using only one reference liquid.
引用
收藏
页码:459 / 462
页数:4
相关论文
共 18 条
[1]  
BAUER H, 1995, PTB-MITT, V105, P99
[2]   PRECISE VISCOSITY MEASUREMENTS OF NEWTONIAN LIQUIDS WITH V LESS-THAN 1 MM2/S FOR THE SELECTION OF SUITABLE STANDARDS [J].
BAUER, H ;
MEERLENDER, G .
RHEOLOGICA ACTA, 1984, 23 (05) :514-521
[3]  
BESTAD DA, 1988, PHYSICA A, V151, P246
[4]  
*DIN, 1983, 515624 DIN 4
[5]  
EINFELDT J, 1982, RHEOL ACTA, V21, P95, DOI 10.1007/BF01520710
[6]  
Einfeldt J., 1970, Rheologica Acta, V9, P424, DOI 10.1007/BF01975411
[7]  
Einfeldt J., 1989, Experimentelle Technik der Physik, V37, P239
[8]  
Einfeldt J., 1983, Experimentelle Technik der Physik, V31, P339
[9]  
EINFELDT J, 1979, Z EXPER PHYS, V27, P271
[10]  
EINFELDT J, 1979, Z EXPER PHYS, V27, P283