Techniques for the detection of coronary atherosclerosis: Multi-detector row CT coronary angiography

被引:144
作者
Vogl, TJ
Abolmaali, ND
Diebold, T
Engelmann, K
Ay, M
Dogan, S
Wimmer-Greinecker, G
Moritz, A
Herzog, C
机构
[1] Univ Frankfurt, Inst Diagnost & Intervent Radiol, D-60590 Frankfurt, Germany
[2] Univ Frankfurt, Dept Cardiovasc Thorac Surg, D-60590 Frankfurt, Germany
关键词
angiography; computed tomography (CT); multidetector row; three-dimensional; coronary vessels; CT; heart;
D O I
10.1148/radiol.2231010515
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
PURPOSE: To investigate the accuracy of different computed tomographic (CT) reformation techniques in assessing the coronary arteries. MATERIALS AND METHODS: Sixty-four patients undergoing both multi-detector row CT and invasive coronary angiography were consecutively included in a retrospective study. CT scans were obtained with collimation of 4 X I mm, pitch of 1.5, and rotation time of 500 msec. Retrospective electrocardiographic gating was used for image reconstruction, with 1.25-mm section thickness and 0.5-mm increment. The CT data set of each patient was evaluated by independent observers using transverse scanning, virtual endoscopic, and three-dimensional reformation and multiplanar reformation. RESULTS: Hemodynamically relevant stenoses (>50%) were detected with highest sensitivity at transverse scanning (58 of 79 [73.4%] stenoses), followed by virtual endoscopic (38 of 79 [48.1 %] stenoses) and three-dimensional reformation (34 of 79 [43.0%] stenoses), and multiplanar reformation (37 of 79 [46.8%] stenoses). Atherosclerotic plaques were identified with comparable sensitivities at transverse scanning (143 of 218 plaques [65.6%]) and at three-dimensional (139 of 218 [63.8%] plaques) and virtual endoscopic reformation (136 of 218 [62.4%] plaques). Multiplanar reformation had distinctly poorer results (217 of 218 [58.3%] plaques). Combined interpretation with all four techniques increased sensitivity to 74.7% (59 of 79) for stenosis and 71.6% (156 of 218) for atherosclerosis. Calculated overall specificity was 91.4% or greater. Sufficient vascular evaluation was possible only in vessels larger than 1.6 mm in diameter. Thus, even in patients with heart rates below 60 bpm, only 80.0% of all coronary segments could be visualized, while at higher frequencies, visibility decreased to 66.2%. CONCLUSION: Although multi-detector row CT is a favorable alternative procedure in evaluating coronary arteries, its clinical value still is restricted to low heart rates and proximal coronary arterial segments. (C) RSNA, 2002.
引用
收藏
页码:212 / 220
页数:9
相关论文
共 29 条
[11]  
FALK E, 1983, BRIT HEART J, V50, P127
[12]   Detection and quantification of coronary calcification: An update. [J].
Fischbach, R ;
Heindel, W .
ROFO-FORTSCHRITTE AUF DEM GEBIET DER RONTGENSTRAHLEN UND DER BILDGEBENDEN VERFAHREN, 2000, 172 (05) :407-414
[13]   Coronary arterial calcium screening: Ready for prime time [J].
Flamm, SD .
RADIOLOGY, 1998, 208 (03) :571-572
[14]   In vitro atherosclerotic plaque and calcium quantitation by intravascular ultrasound and electron-beam computed tomography [J].
Gutfinger, DE ;
Leung, CY ;
Hiro, T ;
Maheswaran, B ;
Nakamura, S ;
Detrano, R ;
Kang, XP ;
Tang, WY ;
Tobis, JM .
AMERICAN HEART JOURNAL, 1996, 131 (05) :899-906
[15]  
Hong C, 2000, RADIOLOGY, V217, P374
[16]  
HONG C, 2000, ROFO FORTSCHR RONTG, P172
[17]   Pathomorphologic classification of coronary atherosclerosis [J].
Ihling, C .
HERZ, 1998, 23 (02) :69-77
[18]   COMPARISON OF SERIAL QUANTITATIVE-EVALUATION OF CALCIFIED CORONARY-ARTERY PLAQUE BY ULTRAFAST COMPUTED-TOMOGRAPHY IN PERSONS WITH AND WITHOUT OBSTRUCTIVE CORONARY-ARTERY DISEASE [J].
JANOWITZ, WR ;
AGATSTON, AS ;
VIAMONTE, M .
AMERICAN JOURNAL OF CARDIOLOGY, 1991, 68 (01) :1-6
[19]   DETECTION OF HEART CALCIFICATION WITH ELECTRON-BEAM CT - INTEROBSERVER AND INTRAOBSERVER RELIABILITY FOR SCORING QUANTIFICATION [J].
KAUFMANN, RB ;
SHEEDY, PF ;
BREEN, JF ;
KELZENBERG, JR ;
KRUGER, BL ;
SCHWARTZ, RS ;
MOLL, PP .
RADIOLOGY, 1994, 190 (02) :347-352
[20]   Cardiac multidetector-row CT:: Retrospectively ECG-gated spiral with optimized temporal and spatial resolution:: First clinical results. [J].
Kopp, AF ;
Ohnesorge, B ;
Flohr, T ;
Georg, C ;
Schröder, S ;
Küttner, A ;
Martensen, J ;
Claussen, CD .
ROFO-FORTSCHRITTE AUF DEM GEBIET DER RONTGENSTRAHLEN UND DER BILDGEBENDEN VERFAHREN, 2000, 172 (05) :429-435