Determination of selenium in urine by inductively coupled plasma mass spectrometry: interferences and optimization

被引:71
作者
Gammelgaard, B [1 ]
Jons, O [1 ]
机构
[1] Royal Danish Sch Pharm, Dept Analyt & Pharmaceut Chem, DK-2100 Copenhagen, Denmark
关键词
D O I
10.1039/a900940j
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
The aim of this study was to develop a method for selenium determination in urine and examine the influence of sensitivity enhancement reagents, instrument parameters, internal standards and the salt content of the urine matrix on the determination. Several carbon-containing solutions (methanol, ethanol, propanol, butanol, glycerol, acetonitrile and acetic acid) were examined for their sensitivity enhancement effect. Enhancement factors up to six were obtained and were dependent on the nebulizer gas flow and rf power. There was no important difference in the enhancement effects of these solutes when the nebulizer gas flow rate was optimized for each solute. The influences of sample uptake rate, nebulizer flow rate and rf power were examined in multivariate experiments. The nebulizer gas flow rate and rf power were found to be interdependent, but the sample pump flow rate was independent of the other parameters. The sensitivities of different selenium species (selenite, selenate, selenomethionine and trimethylselenonium iodide) were equal during the experiments in different enhancement solutes and when analysed with the optimized parameter settings. The influence of the urine matrix and different salts on four selenium isotopes were examined. It was concluded that only Se-82 was usable for quantitative determination in urine as the blank at mass 82 was close to zero. The blank values at masses 76, 77 and 78 varied considerably and differently with different salts and salt concentrations. Yttrium,; indium and gallium were compared for their use as internal standards. There was no difference in the efficiency of these internal standards and they were all usable. The accuracy of the method was determined using the NIST SRM 2670 Toxic Metals in Freeze-dried Urine (certified value 0.030 +/- 0.008 mg l(-1)). The result was 29.4 +/- 1.0 mu g l(-1) (n = 8) and the precision was 3.4%. In synthetic urine diluted 1 + 1, the limit of detection was 0.9 mu g l(-1) and the limit of determination was 2.2 mu g l(-1).
引用
收藏
页码:867 / 874
页数:8
相关论文
共 23 条
[1]   Enzymatic digestion for the determination of trace elements in blood serum by inductively coupled plasma mass spectrometry [J].
AbouShakra, FR ;
Rayman, MP ;
Ward, NI ;
Hotton, V ;
Bastian, G .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1997, 12 (04) :429-433
[2]  
ALAEJOS MS, 1993, CLIN CHEM, V39, P2040
[3]   SIGNAL ENHANCEMENT OF ELEMENTS DUE TO THE PRESENCE OF CARBON-CONTAINING COMPOUNDS IN INDUCTIVELY COUPLED PLASMA MASS-SPECTROMETRY [J].
ALLAIN, P ;
JAUNAULT, L ;
MAURAS, Y ;
MERMET, JM ;
DELAPORTE, T .
ANALYTICAL CHEMISTRY, 1991, 63 (14) :1497-1498
[4]  
BURK RF, 1995, HEPATOLOGY, V21, P561
[5]   Effects of selenium supplementation for cancer prevention in patients with carcinoma of the skin a randomized controlled trial - A randomized controlled trial [J].
Clark, LC ;
Combs, GF ;
Turnbull, BW ;
Slate, EH ;
Chalker, DK ;
Chow, J ;
Davis, LS ;
Glover, RA ;
Graham, GF ;
Gross, EG ;
Krongrad, A ;
Lesher, JL ;
Park, HK ;
Sanders, BB ;
Smith, CL ;
Taylor, JR .
JAMA-JOURNAL OF THE AMERICAN MEDICAL ASSOCIATION, 1996, 276 (24) :1957-1963
[6]   Simple method for the accurate determination of selenium in serum by using inductively coupled plasma mass spectrometry [J].
Delves, HT ;
Sieniawska, CE .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1997, 12 (03) :387-389
[7]   SIMPLE APPROACH TO REDUCING POLYATOMIC ION INTERFERENCES ON ARSENIC AND SELENIUM IN INDUCTIVELY COUPLED PLASMA MASS-SPECTROMETRY [J].
EVANS, EH ;
EBDON, L .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1989, 4 (03) :299-300
[8]   EFFECT OF ORGANIC-SOLVENTS AND MOLECULAR GASES ON POLYATOMIC ION INTERFERENCES IN INDUCTIVELY COUPLED PLASMA MASS-SPECTROMETRY [J].
EVANS, EH ;
EBDON, L .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1990, 5 (06) :425-429
[9]   SYNTHESIS OF [TRIMETHYLSELENONIUM-SE-75 IODIDE FROM [SELENOCYSTINE-SE-75 [J].
FOSTER, SJ ;
GANTHER, HE .
ANALYTICAL BIOCHEMISTRY, 1984, 137 (01) :205-209
[10]   A MATHEMATICAL CORRECTION METHOD FOR SPECTRAL INTERFERENCES ON SELENIUM IN INDUCTIVELY-COUPLED PLASMA-MASS SPECTROMETRY [J].
GOOSSENS, J ;
MOENS, L ;
DAMS, R .
TALANTA, 1994, 41 (02) :187-193