Incidence and implications of abrasion of implantable cardioverter-defibrillator leads

被引:16
作者
DeLurgio, DB
Sathavorn, C
Mera, F
Leon, A
Walter, PF
Langberg, JJ
机构
[1] Emory University Hospital, Atlanta, GA
[2] Emory University Hospital, Atlanta, GA 30322, 1364 Clifton Road, NE
关键词
D O I
10.1016/S0002-9149(97)00153-7
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Severe abrasion of implantable cardioverter-defibrillator leads is frequently found during abdominal generator replacement and occasionally results in lead system failure. Careful inspection of leads at the time of generator replacement will identify such abrasions, and, in some cases, lead repair or replacement may be indicated.
引用
收藏
页码:1409 / &
页数:4
相关论文
共 9 条
  • [1] LONG-TERM COMPLICATIONS OF IMPLANTABLE CARDIOVERTER DEFIBRILLATOR LEAD SYSTEMS
    ALMASSI, GH
    OLINGER, GN
    WETHERBEE, JN
    FEHL, G
    [J]. ANNALS OF THORACIC SURGERY, 1993, 55 (04) : 888 - 892
  • [2] INCIDENCE, PRESENTATION, DIAGNOSIS, AND MANAGEMENT OF MALFUNCTIONING IMPLANTABLE CARDIOVERTER-DEFIBRILLATOR RATE-SENSING LEADS
    DAOUD, EG
    KIRSH, MM
    BOLLING, SF
    STRICKBERGER, SA
    MORADY, F
    KOU, WH
    [J]. AMERICAN HEART JOURNAL, 1994, 128 (05) : 892 - 895
  • [3] CLINICAL-EXPERIENCE, COMPLICATIONS, AND SURVIVAL IN 70 PATIENTS WITH THE AUTOMATIC IMPLANTABLE CARDIOVERTER DEFIBRILLATOR
    ECHT, DS
    ARMSTRONG, K
    SCHMIDT, P
    OYER, PE
    STINSON, EB
    WINKLE, RA
    [J]. CIRCULATION, 1985, 71 (02) : 289 - 296
  • [4] 6-YEAR CLINICAL-EXPERIENCE WITH THE AUTOMATIC IMPLANTABLE CARDIOVERTER DEFIBRILLATOR
    EDEL, TB
    MALONEY, JD
    MOORE, S
    MCGUINN, WP
    GOHN, D
    WILLIAMS, D
    WILKOFF, BL
    MCCARTHY, P
    MORANT, VA
    PEARCE, G
    SIMMONS, TW
    TROHMAN, R
    FIRSTENBERG, M
    CASTLE, LW
    [J]. PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1991, 14 (11): : 1850 - 1854
  • [5] SURVIVAL OF IMPLANTABLE PACEMAKER LEADS
    FURMAN, S
    BENEDEK, ZM
    [J]. PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1990, 13 (12): : 1910 - 1914
  • [6] Incidence of lead system malfunction detected during implantable defibrillator generator replacement
    Goyal, R
    Harvey, M
    Horwood, L
    Bogun, F
    Castellani, M
    Chan, KK
    Daoud, E
    Niebauer, M
    Man, KC
    Morady, F
    Strickberger, SA
    [J]. PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1996, 19 (08): : 1143 - 1146
  • [7] THE AUTOMATIC IMPLANTABLE CARDIOVERTER-DEFIBRILLATOR - EFFICACY, COMPLICATIONS AND SURVIVAL IN PATIENTS WITH MALIGNANT VENTRICULAR ARRHYTHMIAS
    KELLY, PA
    CANNOM, DS
    GARAN, H
    MIRABAL, GS
    HARTHORNE, JW
    HURVITZ, RJ
    VLAHAKES, GJ
    JACOBS, ML
    ILVENTO, JP
    BUCKLEY, MJ
    RUSKIN, JN
    [J]. JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1988, 11 (06) : 1278 - 1286
  • [8] FAILURES OF EPICARDIAL AND TRANSVENOUS LEADS FOR IMPLANTABLE CARDIOVERTER-DEFIBRILLATORS
    MATTKE, S
    MULLER, D
    MARKEWITZ, A
    KAULBACH, H
    SCHMOCKEL, M
    DORWARTH, U
    HOFFMANN, E
    STEINBECK, G
    [J]. AMERICAN HEART JOURNAL, 1995, 130 (05) : 1040 - 1044
  • [9] LONG-TERM OUTCOME WITH THE AUTOMATIC IMPLANTABLE CARDIOVERTER-DEFIBRILLATOR
    WINKLE, RA
    MEAD, RH
    RUDER, MA
    GAUDIANI, VA
    SMITH, NA
    BUCH, WS
    SCHMIDT, P
    SHIPMAN, T
    [J]. JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1989, 13 (06) : 1353 - 1361