Comparison of impedance cardiography with invasive hemodynamic measurements in patients with heart failure secondary to ischemic or nonischemic cardiomyopathy

被引:94
作者
Drazner, MH
Thompson, B
Rosenberg, PB
Kaiser, PA
Boehrer, JD
Baldwin, BJ
Dries, DL
Yancy, CW
机构
[1] Univ Texas, SW Med Ctr, Dept Internal Med,Div Cardiol, Donald W Reynolds Cardiovasc Clin Res Ctr,Heart F, Dallas, TX 75390 USA
[2] St Paul Univ Hosp, Dallas, TX USA
关键词
D O I
10.1016/S0002-9149(02)02257-9
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
In 50 patients with heart failure secondary to ischemic or nonischemic cardiomyopathy, impedance cardiography via BioZ.com was performed simultaneously with right-sided cardiac catheterization. Cardiac output measured with impedance cardiography was significantly correlated with both the thermodilution and Fick cardiac outputs, but the thoracic fluid content assessed by impedance cardiography was not an accurate measure of pulmonary arterial wedge pressure.
引用
收藏
页码:993 / +
页数:4
相关论文
共 6 条
[1]   STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1986, 1 (8476) :307-310
[2]   The effectiveness of right heart catheterization in the initial care of critically ill patients [J].
Connors, AF ;
Speroff, T ;
Dawson, NV ;
Thomas, C ;
Harrell, FE ;
Wagner, D ;
Desbiens, N ;
Goldman, L ;
Wu, AW ;
Califf, RM ;
Fulkerson, WJ ;
Vidaillet, H ;
Broste, S ;
Bellamy, P ;
Lynn, J ;
Knaus, WA .
JAMA-JOURNAL OF THE AMERICAN MEDICAL ASSOCIATION, 1996, 276 (11) :889-897
[3]  
DAVIDSON CJ, 2001, HEART DIS TXB CARDIO, P359
[4]   Tailored therapy using dobutamine and nitroglycerin in advanced heart failure [J].
Drazner, MH ;
Solomon, MA ;
Thompson, B ;
Yancy, CW .
AMERICAN JOURNAL OF CARDIOLOGY, 1999, 84 (08) :941-+
[5]   Right heart catheterization and cardiac complications in patients undergoing noncardiac surgery - An observational study [J].
Polanczyk, CA ;
Rohde, LE ;
Goldman, L ;
Cook, EF ;
Thomas, EJ ;
Marcantonio, ER ;
Mangione, CM ;
Lee, TH .
JAMA-JOURNAL OF THE AMERICAN MEDICAL ASSOCIATION, 2001, 286 (03) :309-314
[6]   Noninvasive assessment of hemodynamics: an emphasis on bioimpedance cardiography [J].
Rosenberg, P ;
Yancy, CW .
CURRENT OPINION IN CARDIOLOGY, 2000, 15 (03) :151-155