Automatic seizure detection in the newborn: methods and initial evaluation

被引:158
作者
Gotman, J
Flanagan, D
Zhang, J
Rosenblatt, B
机构
[1] MCGILL UNIV,MONTREAL,PQ H3A 2B4,CANADA
[2] MONTREAL CHILDRENS HOSP,MONTREAL,PQ H3A 2B4,CANADA
来源
ELECTROENCEPHALOGRAPHY AND CLINICAL NEUROPHYSIOLOGY | 1997年 / 103卷 / 03期
基金
英国医学研究理事会;
关键词
seizure; detection; newborn; EEG;
D O I
10.1016/S0013-4694(97)00003-9
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Seizures are most common in the newborn period, but at that age seizures can be very difficult to identify by clinical observation. Therefore the EEG plays an even greater role in newborns than in older children and adults. The electrographic features of seizures and EEG background in the newborn are, however, very different to those found in adults. We present a set of methods for the automatic detection of seizures in the newborn. The methods are aimed at detecting a wide range of patterns, including rhythmic paroxysmal discharges at a wide range of frequencies, as well as repetitive spike patterns, even when they are not very rhythmic. The methods were developed using EEGs obtained from 55 newborns, recorded at 3 hospitals that used differing monitoring protocols. A total of 281 h of recordings containing 679 seizures were analyzed. An initial evaluation indicated that 71% of the seizures and 78% of seizure clusters (group of seizures separated by less than 90 s) were detected, with a false detection rate of 1.7/h. The methods were developed so that they can be implemented to operate in real time. (C) 1997 Elsevier Science Ireland Ltd.
引用
收藏
页码:356 / 362
页数:7
相关论文
共 21 条
[1]   SPATIAL AND TEMPORAL CHARACTERISTICS OF NEONATAL SEIZURES [J].
BYE, AME ;
FLANAGAN, D .
EPILEPSIA, 1995, 36 (10) :1009-1016
[2]   THE EXACT ICTAL AND INTERICTAL DURATION OF ELECTROENCEPHALOGRAPHIC NEONATAL SEIZURES [J].
CLANCY, RR ;
LEGIDO, A .
EPILEPSIA, 1987, 28 (05) :537-541
[3]  
DREFUSBRISAC C, 1964, NEUROLOGICAL ELECTRO, P250
[4]  
FENICHEL GM, 1984, ELECTROEN CLIN NEURO, V58, P33
[5]   AUTOMATIC RECOGNITION OF INTER-ICTAL EPILEPTIC ACTIVITY IN PROLONGED EEG RECORDINGS [J].
GOTMAN, J ;
IVES, JR ;
GLOOR, P .
ELECTROENCEPHALOGRAPHY AND CLINICAL NEUROPHYSIOLOGY, 1979, 46 (05) :510-520
[6]   AUTOMATIC RECOGNITION OF EPILEPTIC SEIZURES IN THE EEG [J].
GOTMAN, J .
ELECTROENCEPHALOGRAPHY AND CLINICAL NEUROPHYSIOLOGY, 1982, 54 (05) :530-540
[7]   AUTOMATIC SEIZURE DETECTION - IMPROVEMENTS AND EVALUATION [J].
GOTMAN, J .
ELECTROENCEPHALOGRAPHY AND CLINICAL NEUROPHYSIOLOGY, 1990, 76 (04) :317-324
[8]   Evaluation of an automatic seizure detection method for the newborn EEG [J].
Gotman, J ;
Flanagan, D ;
Rosenblatt, B ;
Bye, A ;
Mizrahi, EM .
ELECTROENCEPHALOGRAPHY AND CLINICAL NEUROPHYSIOLOGY, 1997, 103 (03) :363-369
[9]  
LUI A, 1992, ELECTROEN CLIN NEURO, V82, P30
[10]  
MIZRAHI E M, 1986, American Journal of EEG Technology, V26, P81