Sensitive high-Tc SQUID magnetometers for unshielded operation

被引:14
作者
DiIorio, MS [1 ]
Yang, KY
Yoshizumi, S
Haupt, SG
Haran, D
Koch, RH
Milliken, FP
Rozen, JR
Lathrop, DK
Kumar, S
Trammell, HS
机构
[1] Magnesensors Inc, San Diego, CA USA
[2] IBM, TJ Watson Res Ctr, Yorktown Heights, NY USA
[3] Quantum Magnet Inc, San Diego, CA USA
基金
美国国家卫生研究院;
关键词
D O I
10.1109/77.784007
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
We have developed sensitive and highly reliable high-T-c SQUID magnetometers using a reproducible SNS junction fabrication process. In order to enable unshielded operation in the earth's field, we have incorporated flux dams Inco direct-coupled SQUID magnetometers, By using up to four pickup coils in parallel, each with its own flux dam, an effective area of 0.57 mm(2) has been achieved on 24 mm x 24 mm substrates using 100 mu m wide pickup coils. We have demonstrated a magnetic field noise sensitivity of 80 fT/root Hz at 10 Hz and 77 K in an externally looked field of 60 mu T. This magnetic field sensitivity unchanged from the zero-field measurement.
引用
收藏
页码:4428 / 4431
页数:4
相关论文
共 10 条
[1]   Reduction of 1/f noise in high-T-c dc superconducting quantum interference devices cooled in an ambient magnetic field [J].
Dantsker, E ;
Tanaka, S ;
Nilsson, PA ;
Kleiner, R ;
Clarke, J .
APPLIED PHYSICS LETTERS, 1996, 69 (26) :4099-4101
[2]   High-T-c super conducting quantum interference devices with slots or holes: Low 1/f noise in ambient magnetic fields [J].
Dantsker, E ;
Tanaka, S ;
Clarke, J .
APPLIED PHYSICS LETTERS, 1997, 70 (15) :2037-2039
[3]  
DIIORIO MS, 1995, APPL PHYS LETT, V67, P1926
[4]  
DILORIO M, 1998, B AM PHYS SOC, V43, P231
[5]  
DILORIO M, 1997, INT WORKSH SUP ISTEC, P265
[6]   PARAMETER DEPENDENCIES OF CHARACTERISTICS OF A HIGH-T-C DC SUPERCONDUCTING QUANTUM INTERFERENCE DEVICE [J].
ENPUKU, K ;
TOKITA, G ;
MARUO, T ;
MINOTANI, T .
JOURNAL OF APPLIED PHYSICS, 1995, 78 (05) :3498-3503
[7]   FLUX DAM, A METHOD TO REDUCE EXTRA LOW-FREQUENCY NOISE WHEN A SUPERCONDUCTING MAGNETOMETER IS EXPOSED TO A MAGNETIC-FIELD [J].
KOCH, RH ;
SUN, JZ ;
Foglietti, V ;
GALLAGHER, WJ .
APPLIED PHYSICS LETTERS, 1995, 67 (05) :709-711
[8]   Low noise operation of integrated YBa2Cu3O7 magnetometers in static magnetic fields [J].
Krey, S ;
David, B ;
Eckart, R ;
Dossel, O .
APPLIED PHYSICS LETTERS, 1998, 72 (24) :3205-3207
[9]  
LATHROP D, 1997, UNPUB
[10]   Magnetic field-induced noise in directly coupled high T-c superconducting quantum interference device magnetometers [J].
Milliken, FP ;
Brown, SL ;
Koch, RH .
APPLIED PHYSICS LETTERS, 1997, 71 (13) :1857-1859