Thermodynamic temperature determinations of Co-C, Pd-C, Pt-C and Ru-C eutectic fixed-point cells

被引:43
作者
Anhalt, K.
Hartmann, J.
Lowe, D.
Machin, G.
Sadli, M.
Yamada, Y.
机构
[1] Phys Tech Bundesanstalt, D-3300 Braunschweig, Germany
[2] Phys Tech Bundesanstalt, Berlin, Germany
[3] Natl Phys Lab, Teddington TW11 0LW, Middx, England
[4] CNAM, LNE INM, Paris, France
[5] AIST, Natl Metrol Inst Japan, Tsukuba, Ibaraki, Japan
关键词
D O I
10.1088/0026-1394/43/2/S16
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
Thermodynamic temperatures during the melt and the freeze of Co-C, Pd-C, Pt-C and Ru-C metal-carbon fixed-point cells manufactured by LNE-INM/CNAM, NMIJ and NPL were determined by absolutely calibrated filter radiometers traceable to the PTB cryogenic radiometer and a radiance comparison method using an IKE LP3 radiation thermometer. The measurement uncertainties were below 400 mK at temperatures up to 2250 K. The results are in agreement within the combined uncertainties with a study on relative temperature differences of the same set of fixed-point cells. For the fixed-point cells manufactured by NPL the results are compared with a previous thermodynamic temperature measurement.
引用
收藏
页码:S78 / S83
页数:6
相关论文
共 10 条
[1]  
ANHALT K, 2005, P NEWR DAV SWITZ, P289
[2]  
BLOEMBERGEN P, 1999, P TEMPMEKO 99 7 INT, P607
[3]  
Hartmann J, 2002, TEMPMEKO 2001: 8TH INTERNATIONAL SYMPOSIUM ON TEMPERATURE AND THERMAL MEASUREMENT IN INDUSTRY AND SCIENCE, VOL 1 & 2, PROCEEDINGS, P377
[4]  
Hartmann J, 2004, P TEMPMEKO 2004 9 IN, P189
[5]   Reproducible metal-carbon eutectic fixed-points [J].
Lowe, D. ;
Yamada, Y. .
METROLOGIA, 2006, 43 (02) :S135-S139
[6]  
MACHIN G, 2004, P TEMPMEKO 2004 DUBR, P589
[7]  
MACHIN G, 2004, P TEMPMEKO 2004 9 IN, P1049
[8]   Precision blackbody sources for radiometric standards [J].
Sapritsky, VI ;
Khlevnoy, BB ;
Khromchenko, VB ;
Lisiansky, BE ;
Mekhontsev, SN ;
Melenevsky, UA ;
Morozova, SP ;
Prokhorov, AV ;
Samoilov, LN ;
Shapoval, VI ;
Sudarev, KA ;
Zelener, MF .
APPLIED OPTICS, 1997, 36 (22) :5403-5408
[9]   Improved calibration of the spectral responsivity of interference filter radiometers in the visible and near infrared spectral range at PTB [J].
Taubert, DR ;
Friedrich, R ;
Hartmann, J ;
Hollandt, J .
METROLOGIA, 2003, 40 (01) :S35-S38
[10]  
Yamada Y, 2003, AIP CONF PROC, V684, P985, DOI 10.1063/1.1627257