Total Vitamin D Assay Comparison of the Roche Diagnostics "Vitamin D Total" Electrochemiluminescence Protein Binding Assay with the Chromsystems HPLC Method in a Population with both D2 and D3 forms of Vitamin D
被引:55
作者:
Abdel-Wareth, Laila
论文数: 0引用数: 0
h-index: 0
机构:
Sheikh Khalifa Med City, Pathol & Lab Med Inst, Abu Dhabi 51900, U Arab EmiratesSheikh Khalifa Med City, Pathol & Lab Med Inst, Abu Dhabi 51900, U Arab Emirates
Abdel-Wareth, Laila
[1
]
Haq, Afrozul
论文数: 0引用数: 0
h-index: 0
机构:
Sheikh Khalifa Med City, Pathol & Lab Med Inst, Abu Dhabi 51900, U Arab EmiratesSheikh Khalifa Med City, Pathol & Lab Med Inst, Abu Dhabi 51900, U Arab Emirates
Haq, Afrozul
[1
]
Turner, Andrew
论文数: 0引用数: 0
h-index: 0
机构:
Sheikh Khalifa Med City, Pathol & Lab Med Inst, Abu Dhabi 51900, U Arab EmiratesSheikh Khalifa Med City, Pathol & Lab Med Inst, Abu Dhabi 51900, U Arab Emirates
Turner, Andrew
[1
]
Khan, Shoukat
论文数: 0引用数: 0
h-index: 0
机构:
Mil Hosp, Pathol & Lab Med Dept, Riyadh 11159, Saudi ArabiaSheikh Khalifa Med City, Pathol & Lab Med Inst, Abu Dhabi 51900, U Arab Emirates
Khan, Shoukat
[2
]
Salem, Arwa
论文数: 0引用数: 0
h-index: 0
机构:
Sheikh Khalifa Med City, Pathol & Lab Med Inst, Abu Dhabi 51900, U Arab EmiratesSheikh Khalifa Med City, Pathol & Lab Med Inst, Abu Dhabi 51900, U Arab Emirates
Salem, Arwa
[1
]
Mustafa, Faten
论文数: 0引用数: 0
h-index: 0
机构:
Sheikh Khalifa Med City, Pathol & Lab Med Inst, Abu Dhabi 51900, U Arab EmiratesSheikh Khalifa Med City, Pathol & Lab Med Inst, Abu Dhabi 51900, U Arab Emirates
Mustafa, Faten
[1
]
Hussein, Nafiz
论文数: 0引用数: 0
h-index: 0
机构:
Sheikh Khalifa Med City, Pathol & Lab Med Inst, Abu Dhabi 51900, U Arab EmiratesSheikh Khalifa Med City, Pathol & Lab Med Inst, Abu Dhabi 51900, U Arab Emirates
Hussein, Nafiz
[1
]
Pallinalakam, Fasila
论文数: 0引用数: 0
h-index: 0
机构:
Sheikh Khalifa Med City, Pathol & Lab Med Inst, Abu Dhabi 51900, U Arab EmiratesSheikh Khalifa Med City, Pathol & Lab Med Inst, Abu Dhabi 51900, U Arab Emirates
Pallinalakam, Fasila
[1
]
Grundy, Louisa
论文数: 0引用数: 0
h-index: 0
机构:
Sheikh Khalifa Med City, Pathol & Lab Med Inst, Abu Dhabi 51900, U Arab EmiratesSheikh Khalifa Med City, Pathol & Lab Med Inst, Abu Dhabi 51900, U Arab Emirates
Grundy, Louisa
[1
]
Patras, Gemma
论文数: 0引用数: 0
h-index: 0
机构:
Sheikh Khalifa Med City, Pathol & Lab Med Inst, Abu Dhabi 51900, U Arab Emirates
Sheikh Khalifa Med City, Inst Pediat, Abu Dhabi 51900, U Arab EmiratesSheikh Khalifa Med City, Pathol & Lab Med Inst, Abu Dhabi 51900, U Arab Emirates
Patras, Gemma
[1
,3
]
Rajah, Jaishen
论文数: 0引用数: 0
h-index: 0
机构:Sheikh Khalifa Med City, Pathol & Lab Med Inst, Abu Dhabi 51900, U Arab Emirates
Rajah, Jaishen
机构:
[1] Sheikh Khalifa Med City, Pathol & Lab Med Inst, Abu Dhabi 51900, U Arab Emirates
[2] Mil Hosp, Pathol & Lab Med Dept, Riyadh 11159, Saudi Arabia
[3] Sheikh Khalifa Med City, Inst Pediat, Abu Dhabi 51900, U Arab Emirates
25-hydroxyvitamin D;
method evaluation;
electrochemiluminescence;
high performance liquid chromatography;
D DEFICIENCY;
CANCER;
RISK;
D O I:
10.3390/nu5030971
中图分类号:
R15 [营养卫生、食品卫生];
TS201 [基础科学];
学科分类号:
100403 [营养与食品卫生学];
摘要:
This study compared two methods of assaying the 25-hydroxylated metabolites of cholecalciferol (vitamin D3) and ergocalciferol (vitamin D2). A fully automated electrochemiluminescence assay from Roche Diagnostics and an HPLC based method from Chromsystems were used to measure vitamin D levels in surplus sera from 96 individuals, where the majority has the D2 form of the vitamin. Deming regression, concordance rate, correlation and Altman Bland agreement were performed. Seventy two subjects (75%) had a D2 concentration >10 nmol/L while the remaining twenty four subjects had vitamin D2 concentration of less than 10 nmol/L by HPLC. Overall, the Roche Diagnostics method showed a negative bias of -2.59 +/- 4.11 nmol/L on the e602 as compared to the HPLC with a concordance rate of 84%. The concordance rate was 91% in samples with D2 of less than 10 nmol/L and 82% in those with D2 concentration >10 nmol/L. The overall correlation had an r value of 0.77. The r value was higher in samples with D2 levels of less than 10 nmol/L, r = 0.96, as compared to those with D2 values of greater than 10 nmol/L, r = 0.74. The observed bias had little impact on clinical decision and therefore is clinically acceptable.