Quality Standards for Sample Processing, Transportation, and Storage in Hemostasis Testing

被引:103
作者
Funk, Dorothy M. Adcock [1 ]
Lippi, Giuseppe [2 ]
Favaloro, Emmanuel J. [3 ]
机构
[1] Esoterix Inc, Lab & Med Director, Englewood, CO 80112 USA
[2] Acad Hosp, Dept Pathol & Lab Med, Lab Clin Chem & Hematol, Parma, Italy
[3] Westmead Hosp, Inst Clin Pathol & Med Res ICPMR, Dept Hematol, Westmead, NSW, Australia
关键词
preanalytical variability; coagulation testing; prothrombin time; activated partial thromboplastin time; laboratory errors; quality standards; CITRATED WHOLE-BLOOD; PNEUMATIC TUBE TRANSPORT; VON-WILLEBRAND-FACTOR; PLATELET-FUNCTION; PREANALYTICAL VARIABLES; RAPID PREPARATION; PROTHROMBIN TIME; FACTOR-VIII; COAGULATION; PLASMA;
D O I
10.1055/s-0032-1319768
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Samples for hemostasis testing drawn into sodium citrate anticoagulant are vulnerable to the effects of preanalytical variables associated with sample processing, transportation, and storage. These variables include the temperature at which samples are transported and stored; the stability of the samples once processed; whether maintained at room temperature, refrigerated, or frozen; methods of centrifugation; as well as the potential impact of using an automated line. Acknowledgment of these variables, as well as understanding their potential impact on assay results, is imperative to the reporting of high quality and accurate results. This article discusses the preanalytical issues associated with sample processing, transportation, and storage and also presents the ideal conditions for sample handling.
引用
收藏
页码:576 / 585
页数:10
相关论文
共 67 条
[1]   The effect of time and temperature variables on routine coagulation tests [J].
Adcock, D ;
Kressin, D ;
Marlar, RA .
BLOOD COAGULATION & FIBRINOLYSIS, 1998, 9 (06) :463-470
[2]   Minimum specimen volume requirements for routine coagulation testing - Dependence on citrate concentration [J].
Adcock, DM ;
Kressin, DC ;
Marlar, RA .
AMERICAN JOURNAL OF CLINICAL PATHOLOGY, 1998, 109 (05) :595-599
[3]  
Awad Mohammed A, 2004, Hematology, V9, P333, DOI 10.1080/10245330400010646
[4]   Reliability of delayed INR determination: Implications for decentralized anticoagulant care with off-site blood sampling [J].
Baglin, T ;
Luddington, R .
BRITISH JOURNAL OF HAEMATOLOGY, 1997, 96 (03) :431-434
[5]  
Barnes PW., 2002, LAB HEMATOL, V8, P204
[6]   Stability of low molecular weight heparin anti-factor Xa activity in citrated whole blood and plasma [J].
Birri, Nicole ;
Baumgartner, Daniela ;
Conte, Tiziana ;
Huynh, Anh ;
Weller, Katja ;
Pavicic, Marko ;
Reusser, Marianne ;
Siegenthaler, Ilse ;
Barizzi, Gabriela ;
Alberio, Lorenzo .
BRITISH JOURNAL OF HAEMATOLOGY, 2011, 155 (05) :629-631
[7]  
Böhm M, 2006, BLOOD COAGUL FIBRIN, V17, P39
[8]  
Carroll W. E., 2001, Journal of Medicine (Westbury), V32, P83
[9]   Platelet aggregation studies: autologous platelet-poor plasma inhibits platelet aggregation when added to platelet-rich plasma to normalize platelet count [J].
Cattaneo, Marco ;
Lecchi, Anna ;
Zighetti, Maddalena Loredana ;
Lussana, Federico .
HAEMATOLOGICA-THE HEMATOLOGY JOURNAL, 2007, 92 (05) :694-697
[10]   Light Transmission Aggregometry and ATP Release for the Diagnostic Assessment of Platelet Function [J].
Cattaneo, Marco .