Development of a PEDRI free-radical imager using a 0.38 T clinical MRI system

被引:58
作者
Lurie, DJ
Li, HH
Petryakov, S
Zweier, JL
机构
[1] Univ Aberdeen, Dept Biomed Phys & Bioengn, Aberdeen AB25 2ZD, Scotland
[2] Johns Hopkins Univ, Sch Med, EPR Ctr, Baltimore, MD USA
[3] Johns Hopkins Univ, Sch Med, Dept Med, Baltimore, MD 21205 USA
关键词
PEDRI; Overhauser imaging; free radical imaging; low magnetic field; MRI;
D O I
10.1002/mrm.10029
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Proton electron double resonance imaging (PEDRI) uses the Overhauser effect to image the distribution of free-radicals in biological samples and animals. Standard MRI hardware and software is used, with the addition of hardware to irradiate the free-radical-of-interest's EPR resonance. For in vivo applications it must be implemented at a sufficiently low magnetic field to result in an EPR irradiation frequency that will penetrate the sample but will not cause excessive nonresonant power deposition therein. Many clinical MRI systems use resistive magnets that are capable of operating at 10-20 mT, and which could thus be used as PEDRI imagers with the addition of a small amount of extra hardware. This article describes the conversion of a 0.38 T whole-body MRI system for operation as a 20.1 mT small-animal PEDRI imager. The magnet power supply control electronics required a small modification to operate at the lower field strength, but no permanent hardware changes to the MRI console were necessary, and no software modification was required. Frequency down- and up-conversion was used on the NMR RF system, together with a new NMR/EPR dual-resonance RF coil assembly. The system was tested on phantoms containing free-radical solution, and was also used to image the distribution of a free-radical contrast agent injected intravenously into anesthetized mice. Magn Reson Med 47: 181-186, 2002. (C) 2002 Wiley-Liss, Inc.
引用
收藏
页码:181 / 186
页数:6
相关论文
共 14 条
[1]   EFFICIENT DECOUPLER COIL DESIGN WHICH REDUCES HEATING IN CONDUCTIVE SAMPLES IN SUPERCONDUCTING SPECTROMETERS [J].
ALDERMAN, DW ;
GRANT, DM .
JOURNAL OF MAGNETIC RESONANCE, 1979, 36 (03) :447-451
[2]   EPR and DNP properties of certain novel single electron contrast agents intended for oximetric imaging [J].
Ardenkjær-Larsen, JH ;
Laursen, I ;
Leunbach, I ;
Ehnholm, G ;
Wistrand, LG ;
Petersson, JS ;
Golman, K .
JOURNAL OF MAGNETIC RESONANCE, 1998, 133 (01) :1-12
[3]  
Golman K, 1998, ACTA RADIOL, V39, P10
[4]  
Golman K, 2000, J MAGN RESON IMAGING, V12, P929, DOI 10.1002/1522-2586(200012)12:6<929::AID-JMRI17>3.0.CO
[5]  
2-J
[6]   OXIMETRY BY DYNAMIC NUCLEAR-POLARIZATION [J].
GRUCKER, D ;
GUIBERTEAU, T ;
ECLANCHER, B .
MAGNETIC RESONANCE IN MEDICINE, 1995, 34 (02) :219-226
[7]   INVIVO DETECTION OF INJECTED FREE-RADICALS BY OVERHAUSER EFFECT IMAGING [J].
GRUCKER, D .
MAGNETIC RESONANCE IN MEDICINE, 1990, 14 (01) :140-147
[8]   Noninvasive measurement of anatomic structure and intraluminal oxygenation in the gastrointestinal tract of living mice with spatial and spectral EPR imaging [J].
He, GL ;
Shankar, RA ;
Chzhan, M ;
Samouilov, A ;
Kuppusamy, P ;
Zweier, JL .
PROCEEDINGS OF THE NATIONAL ACADEMY OF SCIENCES OF THE UNITED STATES OF AMERICA, 1999, 96 (08) :4586-4591
[9]   ELECTRON SPIN RESONANCE SPECTROMETRY ON WHOLE MOUSE IN-VIVO - OPTIMUM FREQUENCY CONSIDERATIONS [J].
HUTCHISON, JM .
JOURNAL OF PHYSICS E-SCIENTIFIC INSTRUMENTS, 1971, 4 (SEP) :703-+
[10]   High resolution electron paramagnetic resonance imaging of biological samples with a single line paramagnetic label [J].
Kuppusamy, P ;
Wang, PH ;
Chzhan, M ;
Zweier, JL .
MAGNETIC RESONANCE IN MEDICINE, 1997, 37 (04) :479-483