Validation of A&D UA-85X device for blood pressure measurement

被引:11
作者
Bonso, Elisa [1 ]
Ragazzo, Fabio [1 ]
Palatini, Paolo [1 ]
机构
[1] Univ Padua, Dipartimento Med Clin & Sperimentale, I-35128 Padua 2, Italy
关键词
blood pressure; device; hypertension; validation;
D O I
10.1097/MBP.0b013e32830d4b77
中图分类号
R6 [外科学];
学科分类号
1002 ; 100210 ;
摘要
To determine the accuracy of the UA-85X (UA-851, 852, 853, 854, and 855) device developed by the A&D company. Device evaluations were performed using the protocol of the European Society of Hypertension (ESH). Monitor performance was assessed in relation to patients' age, arm circumference, and systolic and diastolic blood pressures (BPs). The device was assessed in two different samples according to ESH requirements, which are based on four zones of accuracy differing from the mercury standard by 5, 10, 15 mmHg, or more. The UA-85X passed all three phases of the protocol for systolic BP and diastolic BR Mean BID difference between device and observers was -0.3 +/- 4.4 mmHg for systolic BP and -2.7 +/- 4.8 mmHg for diastolic Bp. These passed the Association for the Advancement of Medical Instrumentation standard requirements. In multivariable analyses, systolic BP discrepancies between device and observers were related to age (P = 0.03) and diastolic BP discrepancies were related to diastolic BID level (P < 0.001). These data show that the UA-85X satisfies the recommended ESH and Advancement of Medical Instrumentation accuracy levels for both systolic BP and diastolic BP. Blood Press Monit 13:339-341 (C) 2008 Wolters Kluwer Health vertical bar Lippincott Williams & Wilkins.
引用
收藏
页码:339 / 341
页数:3
相关论文
共 5 条
[1]  
[Anonymous], 1987, AM NAT STAND EL AUT
[2]   Consequences of banning mercury and the cuff controversy [J].
O'Brien, E .
BLOOD PRESSURE MONITORING, 2000, 5 (01) :33-34
[3]   Working Group on Blood Pressure Monitoring of the European Society of Hypertension International Protocol for validation of blood pressure measuring devices in adults [J].
O'Brien, E ;
Pickering, T ;
Asmar, R ;
Myers, M ;
Parati, G ;
Staessen, J ;
Mengden, T ;
Imai, Y ;
Waeber, B ;
Palatini, P ;
Atkins, N ;
Gerin, W .
BLOOD PRESSURE MONITORING, 2002, 7 (01) :3-17
[4]   INACCURACY OF 7 POPULAR SPHYGMOMANOMETERS FOR HOME MEASUREMENT OF BLOOD-PRESSURE [J].
OBRIEN, E ;
MEE, F ;
ATKINS, N ;
OMALLEY, K .
JOURNAL OF HYPERTENSION, 1990, 8 (07) :621-634
[5]   British Hypertension Society guidelines for hypertension management 1999: summary [J].
Ramsay, LE ;
Williams, B ;
Johnston, GD ;
MacGregor, GA ;
Poston, L ;
Potter, JF ;
Poulter, NR ;
Russell, G .
BMJ-BRITISH MEDICAL JOURNAL, 1999, 319 (7210) :630-635