Precise measurement of isotope ratios with a double-focusing magnetic sector ICP mass spectrometer

被引:106
作者
Vanhaecke, F
Moens, L
Dams, R
Taylor, P
机构
[1] STATE UNIV GHENT, INST NUCL SCI, ANALYT CHEM LAB, B-9000 GHENT, BELGIUM
[2] COMMISS EUROPEAN COMMUNITIES, JOINT RES CTR, INST REFERENCE MAT & MEASUREMENTS, B-2440 GEEL, BELGIUM
关键词
D O I
10.1021/ac9507247
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
The potential of a commercially available double-focusing magnetic sector ICP mass spectrometer (Element, Finnigan MAT, Bremen, Germany) for precise isotope ratio measurement at the low-resolution setting (R = 300) was evaluated. Optimization of scanning conditions led to a relative standard deviation for a set of 10 consecutive 2 min measurements of similar to 0.1% (Pb-206(+)/Pb-207(+)) at signal intensities of similar to 200 000 counts/s (peak height). This compares favorably with the best values ever reported for quadrupole ICPMS and barely exceeds the theoretical value (counting statistics). increasing the signal intensity to values greater than or equal to 500 000 counts/s (peak height) resulted in a further reduction of the RSDs obtained (for both Mg-25(+)/Mg-26(+) and Pb-206(+)/Pb-207(+)) to typically 0.04%. These figures are remarkably better than those reported for commercially available quadrupole ICPMS systems. This improvement significantly reduces the difference between isotope ratio precision of ICPMS on one hand and those of thermal ionization mass spectrometry and plasma source multiple collector mass spectrometry on the other.
引用
收藏
页码:567 / 569
页数:3
相关论文
共 19 条
[1]   INDUCTIVELY COUPLED PLASMA AS AN ION-SOURCE FOR HIGH-RESOLUTION MASS-SPECTROMETRY [J].
BRADSHAW, N ;
HALL, EFH ;
SANDERSON, NE .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1989, 4 (08) :801-803
[2]   ACCURATE AND PRECISE DETERMINATION OF LEAD ISOTOPE RATIOS IN CLINICAL AND ENVIRONMENTAL-SAMPLES USING INDUCTIVELY COUPLED PLASMA SOURCE-MASS SPECTROMETRY [J].
CAMPBELL, MJ ;
DELVES, HT .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1989, 4 (02) :235-236
[3]   STUDIES IN THE DETERMINATION OF LEAD ISOTOPE RATIOS BY INDUCTIVELY COUPLED PLASMA MASS-SPECTROMETRY [J].
DATE, AR ;
CHEUNG, YY .
ANALYST, 1987, 112 (11) :1531-1540
[4]   OPTIMIZATION OF THE MASS SCANNING RATE FOR THE DETERMINATION OF LEAD ISOTOPE RATIOS USING AN INDUCTIVELY COUPLED PLASMA MASS-SPECTROMETER [J].
FURUTA, N .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1991, 6 (03) :199-203
[5]   LEAD IN ARCHAEOLOGICAL SAMPLES - AN ISOTOPIC STUDY BY ICP-MS [J].
GHAZI, AM .
APPLIED GEOCHEMISTRY, 1994, 9 (06) :627-636
[6]   HIGH-RESOLUTION ICP-MS - A NEW CONCEPT FOR ELEMENTAL MASS-SPECTROMETRY [J].
GIESSMANN, U ;
GREB, U .
FRESENIUS JOURNAL OF ANALYTICAL CHEMISTRY, 1994, 350 (4-5) :186-193
[7]   NOISE SOURCES IN INDUCTIVELY-COUPLED PLASMA-MASS SPECTROMETRY - AN INVESTIGATION OF THEIR IMPORTANCE TO THE PRECISION OF ISOTOPE RATIO MEASUREMENTS [J].
GRAY, AL ;
WILLIAMS, JG ;
INCE, AT ;
LIEZERS, M .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1994, 9 (10) :1179-1184
[8]  
HERTENS RC, 1993, 2ND REGENSBURG S MAS, P2
[9]   INDUCTIVELY COUPLED PLASMA MASS-SPECTROMETRIC DETERMINATION OF LEAD ISOTOPES [J].
HINNERS, TA ;
HEITHMAR, EM ;
SPITTLER, TM ;
HENSHAW, JM .
ANALYTICAL CHEMISTRY, 1987, 59 (22) :2658-2662
[10]   VERIFICATION OF A CORRECTION PROCEDURE FOR MEASUREMENT OF LEAD ISOTOPE RATIOS BY INDUCTIVELY COUPLED PLASMA MASS-SPECTROMETRY [J].
KETTERER, ME ;
PETERS, MJ ;
TISDALE, PJ .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1991, 6 (06) :439-443