Inability of unpacked gooseneck liners to stop the sample liquid after injection with band formation (fast autosampler) into hot GC injectors

被引:7
作者
Bieri, S
Christen, P
Biedermann, M
Grob, K
机构
[1] Univ Geneva, Sch Pharm, Lab Pharmaceut Analyt Chem, CH-1211 Geneva 4, Switzerland
[2] Official Food Control Author Canton Zurich, CH-8030 Zurich, Switzerland
关键词
D O I
10.1021/ac035161a
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
With the introduction of the fast autosampler, the rules on how to perform splitless injection should have been revisited: the fast autosampler releases the sample liquid as a band that is no longer nebulized by solvent vapors as in previous injection techniques. In empty gooseneck liners, the sample liquid is shot to the bottom of the injector and jumps around in a largely uncontrolled manner. Visual experiments showed that the liquid is partially rejected. Another portion passes by the column inlet into the column attachment region, while the last part may directly enter the column. The impact on chromatography was investigated by using a mixture of n-alkanes: the higher boiling components passing by the column entrance into the zone of the column attachment were largely lost, i.e., were discriminated against the volatile components. It was concluded that empty gooseneck liners are not suitable for injection by fast autosamplers.
引用
收藏
页码:1696 / 1701
页数:6
相关论文
共 17 条
[1]  
BIEDERMANN M, 2000, VISUALIZATION EVAPOR
[2]  
GROB K, 1992, HRC-J HIGH RES CHROM, V15, P190
[3]   Video-taped sample evaporation in hot chambers simulating gas chromatography split/splitless injectors - I. Thermospray injection [J].
Grob, K ;
Biedermann, M .
JOURNAL OF CHROMATOGRAPHY A, 2000, 897 (1-2) :237-246
[4]  
GROB K, 1992, HRC-J HIGH RES CHROM, V15, P399
[5]   SHOULD THE SEPTUM PART OF VAPORIZING INJECTORS BE KEPT AT LOWER TEMPERATURES [J].
GROB, K ;
NEUKOM, HP .
JOURNAL OF CHROMATOGRAPHY, 1980, 198 (01) :64-69
[6]  
GROB K, 1992, HRC-J HIGH RES CHROM, V15, P335
[7]   Video-taped sample evaporation in hot chambers simulating gas chromatography split and splitless injectors - II. Injection with band formation [J].
Grob, K ;
Biedermann, M .
JOURNAL OF CHROMATOGRAPHY A, 2000, 897 (1-2) :247-258
[8]  
GROB K, 1993, HRC-J HIGH RES CHROM, V16, P464
[9]   SAMPLE TRANSFER IN SPLITLESS INJECTIONS IN CAPILLARY GAS-CHROMATOGRAPHY [J].
GROB, K ;
ROMANN, A .
JOURNAL OF CHROMATOGRAPHY, 1981, 214 (01) :118-121
[10]   The two options for sample evaporation in hot GC injectors: Thermospray and band formation. Optimization of conditions and injector design [J].
Grob, K ;
Biedermann, M .
ANALYTICAL CHEMISTRY, 2002, 74 (01) :10-16