Comparison of measurement accuracy between two types of wrist goniometer systems

被引:48
作者
Jonsson, P
Johnson, PW
机构
[1] Sahlgrens Univ Hosp, Environm & Occupat Med Sect, SE-41266 Gothenburg, Sweden
[2] Univ Washington, Dept Environm Hlth, Seattle, WA 98195 USA
关键词
wrist; goniometer and accuracy;
D O I
10.1016/S0003-6870(01)00036-9
中图分类号
T [工业技术];
学科分类号
08 ;
摘要
Studies have shown that wrist goniometers are prone to measurement errors, particularly due to crosstalk. This study compared two wrist goniometer systems: a commonly used biaxial, single transducer (System A) and a biaxial, two-transducer (System B). Wrist angles, range of movement and crosstalk results were compared. With the wrist in 90 degrees of pronation, eight subjects were placed in 20 different wrist postures between -40 degrees and 40 degrees of flexion/extension and between -10 degrees and 20 degrees of deviation. Relative to System B, System A had larger measurement errors and was more prone to crosstalk. There may be two sources of crosstalk: (1) intrinsic crosstalk associated with the design, application and twisting of the goniometer transducer when on the wrist, and (2) extrinsic crosstalk associated with the anatomy and complex movement of the wrist joint. It appears that the majority of the radial/ulnar crosstalk measured with System A was intrinsic crosstalk due to the twisting of the goniometer transducer. (C) 2001 Elsevier Science Ltd. All rights reserved.
引用
收藏
页码:599 / 607
页数:9
相关论文
共 19 条
[1]  
[Anonymous], ADV OCCUP MED REHABI
[2]   Practical operation of a biaxial goniometer at the wrist joint [J].
Buchholz, B ;
Wellman, H .
HUMAN FACTORS, 1997, 39 (01) :119-129
[3]  
EDWARDS GL, 1996, Patent No. 5533531
[4]  
Garrett J. W, 1970, AMRLTR6942 WRIGHT PA
[5]  
GARRETT JW, 1970, AMRLTR6926 WRIGHT PA
[6]  
Greene WB., 1994, The Clinical Measurement of Joint Motion
[7]  
HABBERG M, 1995, WORK RELATED MUSCULO
[8]   Forearm muscular load and wrist angle among automobile assembly line workers in relation to symptoms [J].
Hagg, GM ;
Oster, J ;
Bystrom, S .
APPLIED ERGONOMICS, 1997, 28 (01) :41-47
[9]   Goniometer measurement and computer analysis of wrist angles and movements applied to occupational repetitive work [J].
Hansson, GA ;
Balogh, I ;
Ohlsson, K ;
Rylander, L ;
Skerfving, S .
JOURNAL OF ELECTROMYOGRAPHY AND KINESIOLOGY, 1996, 6 (01) :23-35
[10]   WRIST MOTIONS IN INDUSTRY [J].
MARRAS, WS ;
SCHOENMARKLIN, RW .
ERGONOMICS, 1993, 36 (04) :341-351