Adjustment of serum total calcium for albumin concentration: Effects of non-linearity and of regression differences between laboratories

被引:36
作者
Barth, JH
Fiddy, JB
Payne, RB
机构
[1] YORK DIST GEN HOSP,DEPT CHEM PATHOL,YORK,N YORKSHIRE,ENGLAND
[2] ST JAMES UNIV HOSP,DEPT CHEM PATHOL,LEEDS,W YORKSHIRE,ENGLAND
关键词
reference ranges; probit analysis;
D O I
10.1177/000456329603300108
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
It has been reported that the relationship between serum total calcium and albumin concentrations in hospital patients deviates from linearity at low albumin concentrations. We searched a large laboratory computer data base for adult patients with discretionary requests for serum calcium analysis but with no other data suggestive of disturbances of calcium homeostasis, and collected a minimum of 100 calcium values at each of a wide range of albumin concentrations. We confirmed the earlier observation, but found that the use of a single regression to derive an adjustment of total calcium for albumin gave only small differences of no clinical significance. To investigate whether equations to adjust total calcium for albumin can be transferred between laboratories, three laboratory computers were searched for calcium requests from patients likely to have a low prevalence of calcium disturbances. The regressions of total calcium on albumin differed significantly between laboratories, but within each laboratory gave adjusted calcium values identical with those in health. Although the errors resulting from applying an equation from one laboratory to another were small and unlikely to be of major clinical significance, we recommend that where possible laboratories should derive adjustment equations from their own data.
引用
收藏
页码:55 / 58
页数:4
相关论文
共 8 条
[1]   THE ADJUSTED SERUM-CALCIUM CONCEPT - A REAPPRAISAL [J].
ASHBY, JP ;
WRIGHT, DJ ;
RINSLER, MG .
ANNALS OF CLINICAL BIOCHEMISTRY, 1986, 23 :533-537
[2]   THE MEASUREMENT OF ALBUMIN IN SERUM AND PLASMA [J].
HILL, PG .
ANNALS OF CLINICAL BIOCHEMISTRY, 1985, 22 (NOV) :565-578
[3]  
MASTERS PW, 1993, CLIN CHEM, V39, P1082
[4]   SERUM-ALBUMIN BY DYE-BINDING - BROMOCRESOL GREEN OR BROMOCRESOL PURPLE - THE CASE FOR CONSERVATISM [J].
MCGINLAY, JM ;
PAYNE, RB .
ANNALS OF CLINICAL BIOCHEMISTRY, 1988, 25 :417-421
[5]   PROTEIN INTERFERES WITH IONIZED CALCIUM MEASUREMENT AT THE REFERENCE ELECTRODE LIQUID JUNCTION [J].
PAYNE, RB ;
JONES, DP .
ANNALS OF CLINICAL BIOCHEMISTRY, 1987, 24 :400-407
[6]   INTERPRETATION OF SERUM-CALCIUM IN PATIENTS WITH ABNORMAL SERUM-PROTEINS [J].
PAYNE, RB ;
LITTLE, AJ ;
WILLIAMS, RB ;
MILNER, JR .
BMJ-BRITISH MEDICAL JOURNAL, 1973, 4 (5893) :643-646
[7]   INTERPRETATION OF SERUM TOTAL CALCIUM - EFFECTS OF ADJUSTMENT FOR ALBUMIN CONCENTRATION ON FREQUENCY OF ABNORMAL VALUES AND ON DETECTION OF CHANGE IN THE INDIVIDUAL [J].
PAYNE, RB ;
CARVER, ME ;
MORGAN, DB .
JOURNAL OF CLINICAL PATHOLOGY, 1979, 32 (01) :56-60