Clinical experience in end-to-side venous anastomoses with a microvascular anastomotic coupling device in head and neck reconstruction

被引:49
作者
DeLacure, MD
Kuriakose, MA
Spies, AL
机构
[1] NYU, Sch Med, Dept Otolaryngol, New York, NY 10016 USA
[2] Roswell Pk Canc Inst, Dept Head & Neck Surg & Oncol, Div Plast & Reconstruct Surg, Buffalo, NY USA
关键词
D O I
10.1001/archotol.125.8.869
中图分类号
R76 [耳鼻咽喉科学];
学科分类号
100213 [耳鼻咽喉科学];
摘要
Background: Microvascular anastomosis remains one of the most technically sensitive aspects of free-tissue transfer reconstructions. Despite the availability of various mechanical anastomotic coupling systems for human clinical use during the last 8 years, reported clinical series remain rare. Objective: To describe a clinical experience in applying a mechanical microvascular anastomotic coupling device (MACD) to end-to-side anastomotic configurations in head and neck free-flap reconstruction. Methods: The MACD is a readily available high-density polyethylene ring-stainless steel pin system that has been found to be highly effective in clinical studies of end-to-end arterial and venous anastomosis and in laboratory studies of end-to-side anastomosis of rabbit arteries. Results: Thirty-seven end-to-side venous anastomoses were attempted, of which 33 (89%) were completed. Of these, 9 patients had critical anastomoses (only 1 venous anastomosis per patient). In patients undergoing parallel venous anastomoses, 6 had both anastomoses performed using the MACD; in the remaining 12 patients, 1 of the anastomoses was performed using the MACD. A variety of donor flaps and clinical contexts were encountered. Flap survival in the MACD series was 100%. Four anastomoses were converted to conventional suture technique intraoperatively. Conclusions: The MACD is well suited to end-to-side venous anastomosis when carefully and selectively used by experienced microvascular surgeons. The most common clinical, situation requiring this configuration and technique was the lack of ipsilateral recipient veins for end-to-end anastomosis or a ligated internal jugular vein stump that required this approach for device application. Previous radiation therapy does not appear to be a contraindication to its use.
引用
收藏
页码:869 / 872
页数:4
相关论文
共 17 条
[1]
CLINICAL-EXPERIENCE WITH THE 3M MICROVASCULAR COUPLING ANASTOMOTIC DEVICE IN 100 FREE-TISSUE TRANSFERS [J].
AHN, CY ;
SHAW, WW ;
BERNS, S ;
MARKOWITZ, BL .
PLASTIC AND RECONSTRUCTIVE SURGERY, 1994, 93 (07) :1481-1484
[2]
A MICROVASCULAR ANASTOMOTIC DEVICE .1. A HEMODYNAMIC EVALUATION IN RABBIT FEMORAL ARTERIES AND VEINS [J].
BLAIR, WF ;
STEYERS, CM ;
BROWN, TD ;
GABLE, RH .
MICROSURGERY, 1989, 10 (01) :21-28
[3]
A MICROVASCULAR ANASTOMOTIC DEVICE .2. A HISTOLOGIC-STUDY IN ARTERIES AND VEINS [J].
BLAIR, WF ;
MORECRAFT, RJ ;
STEYERS, CM ;
MAYNARD, JA .
MICROSURGERY, 1989, 10 (01) :29-39
[4]
DAUTEL G, 1990, 6 ANN M AM SOC REC M
[5]
CLINICAL-EXPERIENCE WITH A MICROVASCULAR ANASTOMOTIC DEVICE IN HEAD AND NECK RECONSTRUCTION [J].
DELACURE, MD ;
WONG, RS ;
MARKOWITZ, BL ;
KOBAYASHI, MR ;
AHN, CY ;
SHEDD, DP ;
SPIES, AL ;
LOREE, TR ;
SHAW, WW .
AMERICAN JOURNAL OF SURGERY, 1995, 170 (05) :521-523
[6]
FALCONER DP, 1996, J RECONSTR MICROSURG, V6, P215
[7]
STRENGTH OF MICROVASCULAR ANASTOMOSES - COMPARISON BETWEEN THE UNILINK ANASTOMOTIC SYSTEM AND SUTURES [J].
GILBERT, RW ;
RAGNARSSON, R ;
BERGGREN, A ;
OSTRUP, L .
MICROSURGERY, 1989, 10 (01) :40-46
[8]
GILBERT RW, 1989, ARCH OTOLARYNGOL, V115, P970
[9]
NAKAYAMA K, 1962, SURGERY, V52, P918
[10]
THE UNILINK SYSTEM FOR MECHANICAL MICROVASCULAR ANASTOMOSIS IN HAND SURGERY [J].
NYLANDER, G ;
RAGNARSSON, R ;
BERGGREN, A ;
OSTRUP, LT .
JOURNAL OF HAND SURGERY-AMERICAN VOLUME, 1989, 14A (01) :44-48