We have previously reported the short-term migration of cemented Hinek femoral components using radiostereometric analysis (RSA). We now report the mid-term migration. During the first 2 years after implantation the prosthesis subsided into varus and rotated internally. Between years 3 and 8 the prosthesis continued to rotate internally with the head moving posteriorly (0.07 mm/year, P=0.004). It also continued to fall into varus with the tip moving laterally (0.07 mm/year, P=0.04). The head (0.06 mm/year, P<0.0001), shoulder (0.04 mm/year, P=0.0001) and tip (0.04 mm/year, P=0.001) continued to migrate distally. There were two cases of failure due to aseptic loosening during the follow-up period. During the second year both of these had posterior head migration, which was abnormally rapid (>2 SD from the mean). We have demonstrated that a cemented implant has slow but significant: levels of migration and rotation for at least 8 years after implantation. Our study confirms that implants with abnormally rapid posterior bead migration during the second year are likely to fail.
引用
收藏
页码:140 / 144
页数:5
相关论文
共 14 条
[1]   Cement migration after THR -: A comparison of Charnley Elite and Exeter femoral stems using RSA [J].
Alfaro-Adrián, J ;
Gill, HS ;
Murray, DW .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1999, 81B (01) :130-134
[2]  
ALFAROADRIAN J, 1998, J BONE JOINT SURG BR, V80, P270
[3]   EARLY MIGRATION AND LATE ASEPTIC FAILURE OF PROXIMAL FEMORAL PROSTHESES [J].
FREEMAN, MAR ;
PLANTEBORDENEUVE, P .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1994, 76B (03) :432-438
[4]   WHY RESECT THE NECK [J].
FREEMAN, MAR .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1986, 68 (03) :346-349
[5]  
Gill H S, 1998, J BIOMECH, V31, P88
[6]  
KARRHOLM J, 1993, CLIN ORTHOP RELAT R, P50
[7]   DOES EARLY MICROMOTION OF FEMORAL STEM PROSTHESES MATTER - 4-7-YEAR STEREORADIOGRAPHIC FOLLOW-UP OF 84 CEMENTED PROSTHESES [J].
KARRHOLM, J ;
BORSSEN, B ;
LOWENHIELM, G ;
SNORRASON, F .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1994, 76B (06) :912-917
[8]   Migration of cemented femoral components after THR - Roentgen stereophotogrammetric analysis [J].
Kiss, J ;
Murray, DW ;
TurnerSmith, AR ;
Bithell, J ;
Bulstrode, CJ .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1996, 78B (05) :796-801
[9]  
Kiss J, 1995, Proc Inst Mech Eng H, V209, P169, DOI 10.1243/PIME_PROC_1995_209_340_02
[10]   INSTABILITY, MIGRATION AND LAXITY OF TOTAL HIP PROSTHESES - A ROENTGEN STEREOPHOTOGRAMMETRIC STUDY [J].
MJOBERG, B ;
HANSSON, LI ;
SELVIK, G .
ACTA ORTHOPAEDICA SCANDINAVICA, 1984, 55 (02) :141-145