Development of an optimized dextran retention test for large pore size hollow fiber ultrafiltration membranes

被引:17
作者
Bakhshayeshi, Meisam [1 ]
Teella, Achyuta [1 ]
Zhou, Hongyi [2 ]
Olsen, Cathryn [2 ]
Yuan, Wei [3 ]
Zydney, Andrew L. [1 ]
机构
[1] Penn State Univ, Dept Chem Engn, University Pk, PA 16802 USA
[2] GE Global Res, Membrane & Separat Technol, Niskayuna, NY 12309 USA
[3] GE Healthcare Life Sci, Filtrat R&D, Westborough, MA 01581 USA
关键词
Dextran; Membrane characterization; Hollow fiber; Concentration polarization; Bioprocessing; Molecular weight cut-off;
D O I
10.1016/j.memsci.2012.06.020
中图分类号
TQ [化学工业];
学科分类号
081705 [工业催化];
摘要
Although dextran retention tests are used extensively for characterizing ultrafiltration membranes, there is no standardization in the test methods or conditions. This is particularly important for hollow fiber membrane modules in which the degree of concentration polarization varies with axial position and there can be significant backfiltration when the device is operated at very low permeate flow rates. We have developed a theoretical framework that examines the effects of module geometry (number of fibers, fiber diameter, and fiber length) and operating conditions (feed and permeate flow rates) on the key factors governing the dextran retention profile in hollow fiber membrane modules. This framework was then used to identify optimal conditions for performing dextran retention tests in hollow fiber ultrafiltration membranes with relatively large pore size. The measured dextran retention coefficients at these optimal conditions were in good agreement with model calculations. In addition, the dextran retention profiles under these optimal conditions were essentially insensitive to variations in the feed and permeate flow rate, providing a robust dextran retention test that can yield important information on the membrane pore size and transport characteristics. (C) 2012 Elsevier B.V. All rights reserved.
引用
收藏
页码:32 / 38
页数:7
相关论文
共 9 条
[1]
Understanding dextran retention data for hollow fiber ultrafiltration membranes [J].
Bakhshayeshi, Meisam ;
Zhou, Hongyi ;
Olsen, Cathryn ;
Yuan, Wei ;
Zydney, Andrew L. .
JOURNAL OF MEMBRANE SCIENCE, 2011, 385 (1-2) :243-250
[2]
Sieving characteristics of multilayer ultrafiltration membranes [J].
Boyd, RF ;
Zydney, AL .
JOURNAL OF MEMBRANE SCIENCE, 1997, 131 (1-2) :155-165
[3]
Improvement of a method for the characterization of ultrafiltration membranes by measurements of tracers retention [J].
Causserand, C ;
Rouaix, S ;
Akbari, A ;
Aimar, P .
JOURNAL OF MEMBRANE SCIENCE, 2004, 238 (1-2) :177-190
[4]
Characterization of ultrafiltration membranes by tracer's retention: Comparison of methods sensitivity and reproducibility [J].
Causserand, Christel ;
Pierre, Gwenaelle ;
Rapenne, Sophie ;
Schrotter, Jean-Christophe ;
Sauvade, Patrick ;
Lorain, Olivier .
DESALINATION, 2010, 250 (02) :767-772
[5]
A REJECTION PROFILE TEST FOR ULTRAFILTRATION MEMBRANES AND DEVICES [J].
TKACIK, G ;
MICHAELS, S .
BIO-TECHNOLOGY, 1991, 9 (10) :941-+
[6]
Membrane ultrafiltration in hollow-fiber module with the consideration of pressure declination along the fibers [J].
Yeh, HM ;
Cheng, TW ;
Wu, HH .
SEPARATION AND PURIFICATION TECHNOLOGY, 1998, 13 (03) :171-180
[7]
Zeman L.J., 1996, MICROFILTRATION ULTR
[8]
USE OF THE LOG-NORMAL PROBABILITY DENSITY-FUNCTION TO ANALYZE MEMBRANE PORE-SIZE DISTRIBUTIONS - FUNCTIONAL FORMS AND DISCREPANCIES [J].
ZYDNEY, AL ;
AIMAR, P ;
MEIRELES, M ;
PIMBLEY, JM ;
BELFORT, G .
JOURNAL OF MEMBRANE SCIENCE, 1994, 91 (03) :293-298
[9]
Improving dextran tests for ultrafiltration membranes: Effect of device format [J].
Zydney, Andrew L. ;
Xenopoulos, Alex .
JOURNAL OF MEMBRANE SCIENCE, 2007, 291 (1-2) :180-190