The Bionic Glove: An electrical stimulator garment that provides controlled grasp and hand opening in quadriplegia

被引:132
作者
Prochazka, A
Gauthier, M
Wieler, M
Kenwell, Z
机构
[1] Division of Neuroscience, University of Alberta, Edmonton, Alta.
[2] Division of Neuroscience, 507 HMRC, University of Alberta, Edmonton
来源
ARCHIVES OF PHYSICAL MEDICINE AND REHABILITATION | 1997年 / 78卷 / 06期
基金
英国医学研究理事会;
关键词
D O I
10.1016/S0003-9993(97)90426-3
中图分类号
R49 [康复医学];
学科分类号
100215 ;
摘要
Objective: This report describes the operation of the Bionic Glove, a new functional electrical stimulation (FES) device designed to improve the function of the paralyzed hand after spinal cord injury (SCI) or stroke. Design: Signals from a sensor in the glove detecting voluntary wrist movement are used to control FES of muscles either to produce hand-grasp or to open the hand. When the glove is donned, conductive areas on its inside surface automatically make contact with self-adhesive electrodes on the skin. Setting and Patients: This report concerns nine people with SCI who have used the device in their daily lives for up to a year or more. Measurements were made at clinics in Edmonton, Miami, and Chicago as part of a multicenter clinical trial. Outcome Measures and Results: The mean peak force of tenodesis grasp in the nine subjects increased from 2.6N (passive) to 11.3N (glove active). Active force was significantly greater than passive grasp force even when muscles were fatigued after repetitive grasp-release cycles. Most manual tasks improved significantly with the use of the glove, as judged by the number of tasks completed in a minute or the subjects' qualitative ratings of task difficulty. Conclusion: The Bionic Glove can provide significant improvement of hand function in people with C6-C7 SCI. (C) 1997 by the American Congress of Rehabilitation Medicine and the American Academy of Physical Medicine and Rehabilitation.
引用
收藏
页码:608 / 614
页数:7
相关论文
共 33 条
[1]   ELECTRICAL-STIMULATION OF WRIST AND FINGERS FOR HEMIPLEGIC PATIENTS [J].
BAKER, LL ;
YEH, C ;
WILSON, D ;
WATERS, RL .
PHYSICAL THERAPY, 1979, 59 (12) :1495-1499
[2]   MAMMALIAN MOTOR UNITS - PHYSIOLOGICAL-HISTOCHEMICAL CORRELATION IN 3 TYPES IN CAT GASTROCNEMIUS [J].
BURKE, RE ;
LEVINE, DN ;
ZAJAC, FE ;
TSAIRIS, P ;
ENGEL, WK .
SCIENCE, 1971, 174 (4010) :709-&
[3]  
Hines A. E., 1995, IEEE Transactions on Rehabilitation Engineering, V3, P193, DOI 10.1109/86.392368
[4]   ATTENUATION OF PATHOLOGICAL TREMORS BY FUNCTIONAL ELECTRICAL-STIMULATION .2. CLINICAL-EVALUATION [J].
JAVIDAN, M ;
ELEK, J ;
PROCHAZKA, A .
ANNALS OF BIOMEDICAL ENGINEERING, 1992, 20 (02) :225-236
[5]  
KILGORE KL, 1996, 1 ANN C INT FES SOC
[6]  
Kralj A.R., 1989, Functional Electrical Stimulation: Standing and Walking after Spinal Cord Injury
[7]  
LIBERSON W T, 1961, Arch Phys Med Rehabil, V42, P101
[8]   NEURAL PROSTHETIC INTERFACES WITH THE NERVOUS-SYSTEM [J].
LOEB, GE .
TRENDS IN NEUROSCIENCES, 1989, 12 (05) :195-201
[9]  
LOEB GE, 1991, MED BIOL ENG COMPUT, V29, P13
[10]  