Outpatient Safety Assessment of an In-Home Predictive Low-Glucose Suspend System with Type 1 Diabetes Subjects at Elevated Risk of Nocturnal Hypoglycemia

被引:71
作者
Buckingham, Bruce A. [1 ,3 ]
Cameron, Fraser [2 ]
Calhoun, Peter [3 ]
Maahs, David M. [4 ]
Wilson, Darrell M. [1 ]
Chase, H. Peter [4 ]
Bequette, B. Wayne [2 ]
Lum, John [3 ]
Sibayan, Judy [3 ]
Beck, Roy W. [3 ]
Kollman, Craig [3 ]
机构
[1] Stanford Univ, Stanford, CA 94305 USA
[2] Rensselaer Polytech Inst, Troy, NY USA
[3] Jaeb Ctr Hlth Res, Tampa, FL 33647 USA
[4] Univ Colorado, Barbara Davis Ctr Childhood, Denver, CO 80202 USA
关键词
PUMP THERAPY; INSULIN; PREVENTION; REDUCTION; DURATION; CHILDREN; SENSOR;
D O I
10.1089/dia.2013.0040
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Objective: Nocturnal hypoglycemia is a common problem with type 1 diabetes. In the home setting, we conducted a pilot study to evaluate the safety of a system consisting of an insulin pump and continuous glucose monitor communicating wirelessly with a bedside computer running an algorithm that temporarily suspends insulin delivery when hypoglycemia is predicted. Research Design and Methods: After the run-in phase, a 21-night randomized trial was conducted in which each night was randomly assigned 2:1 to have either the predictive low-glucose suspend (PLGS) system active (intervention night) or inactive (control night). Three predictive algorithm versions were studied sequentially during the study for a total of 252 intervention and 123 control nights. The trial included 19 participants 18-56 years old with type 1 diabetes (hemoglobin A1c level of 6.0-7.7%) who were current users of the MiniMed Paradigm (R) REAL-Time Revel (TM) System and Sof-sensor (R) glucose sensor (Medtronic Diabetes, Northridge, CA). Results: With the final algorithm, pump suspension occurred on 53% of 77 intervention nights. Mean morning glucose level was 144 +/- 48 mg/dL on the 77 intervention nights versus 133 +/- 57 mg/dL on the 37 control nights, with morning blood ketones >0.6 mmol/L following one intervention night. Overnight hypoglycemia was lower on intervention than control nights, with at least one value <= 70 mg/dL occurring on 16% versus 30% of nights, respectively, with the final algorithm. Conclusions: This study demonstrated that the PLGS system in the home setting is safe and feasible. The preliminary efficacy data appear promising with the final algorithm reducing nocturnal hypoglycemia by almost 50%.
引用
收藏
页码:622 / 627
页数:6
相关论文
共 14 条
[1]  
Agrawal Pratik, 2011, J Diabetes Sci Technol, V5, P1137
[3]   Duration of Nocturnal Hypoglycemia Before Seizures [J].
Buckingham, Bruce ;
Wilson, Darrell M. ;
Lecher, Todd ;
Hanas, Ragnar ;
Kaiserman, Kevin ;
Cameron, Fergus .
DIABETES CARE, 2008, 31 (11) :2110-2112
[4]  
Buckingham Bruce, 2005, Diabetes Technol Ther, V7, P440, DOI 10.1089/dia.2005.7.440
[5]   Prevention of Nocturnal Hypoglycemia Using Predictive Alarm Algorithms and Insulin Pump Suspension [J].
Buckingham, Bruce ;
Chase, H. Peter ;
Dassau, Eyal ;
Cobry, Erin ;
Clinton, Paula ;
Gage, Victoria ;
Caswell, Kimberly ;
Wilkinson, John ;
Cameron, Fraser ;
Lee, Hyunjin ;
Bequette, B. Wayne ;
Doyle, Francis J., III .
DIABETES CARE, 2010, 33 (05) :1013-1017
[6]  
Cameron Fraser, 2012, J Diabetes Sci Technol, V6, P1142
[7]   Insulin Pump Therapy With Automated Insulin Suspension in Response to Hypoglycemia Reduction in nocturnal hypoglycemia in those at greatest risk [J].
Choudhary, Pratik ;
Shin, John ;
Wang, Yongyin ;
Evans, Mark L. ;
Hammond, Peter J. ;
Kerr, David ;
Shaw, James A. M. ;
Pickup, John C. ;
Amiel, Stephanie A. .
DIABETES CARE, 2011, 34 (09) :2023-2025
[8]   Prevention of Hypoglycemia by Using Low Glucose Suspend Function in Sensor-Augmented Pump Therapy [J].
Danne, Thomas ;
Kordonouri, Olga ;
Holder, Martin ;
Haberland, Holger ;
Golembowski, Sven ;
Remus, Kerstin ;
Blaesig, Sara ;
Wadien, Tanja ;
Zierow, Susanne ;
Hartmann, Reinhard ;
Thomas, Andreas .
DIABETES TECHNOLOGY & THERAPEUTICS, 2011, 13 (11) :1129-1134
[9]   Reduction in Duration of Hypoglycemia by Automatic Suspension of Insulin Delivery: The In-Clinic ASPIRE Study [J].
Garg, Satish ;
Brazg, Ronald L. ;
Bailey, Timothy S. ;
Buckingham, Bruce A. ;
Slover, Robert H. ;
Klonoff, David C. ;
Shin, John ;
Welsh, John B. ;
Kaufman, Francine R. .
DIABETES TECHNOLOGY & THERAPEUTICS, 2012, 14 (03) :205-209
[10]   Accuracy of an electrochemical sensor for measuring capillary blood ketones by fingerstick samples during metabolic deterioration after continuous subcutaneous insulin infusion interruption in type 1 diabetic patients [J].
Guerci, B ;
Benichou, M ;
Floriot, M ;
Bohme, P ;
Fougnot, S ;
Franck, P ;
Drouin, P .
DIABETES CARE, 2003, 26 (04) :1137-1141