Niobium/molybdenum K-edge filtration in mammography: Contrast and dose evaluation

被引:13
作者
Calicchia, A
Gambaccini, M
Indovina, PL
Mazzei, F
Pugliani, L
机构
[1] UNIV FERRARA,DIPARTMENTO FIS,I-44100 FERRARA,ITALY
[2] IST NAZL FIS NUCL,SEZ FERRARA,FERRARA,ITALY
[3] UNIV NAPLES,DIPARTIMENTO SCI FIS,NAPLES,ITALY
[4] IST NAZL FIS NUCL,SEZ NAPOLI,I-80125 NAPLES,ITALY
[5] IST NAZL FIS NUCL,SEZ SANITA,I-00161 ROME,ITALY
关键词
D O I
10.1088/0031-9155/41/9/011
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The use of a dual Kedge filter (niobium-molybdenum) with a Mo anode x-ray tube for application to mammography is investigated. The incident and transmitted energy spectral distributions are compared with those provided by a molybdenum anode molybdenum filter tube (standard source). The imaging characteristics in terms of contrast and mean glandular dose have been evaluated for various phantom thicknesses. The niobium filtration removes the molybdenum K beta line almost completely from the beam spectrum whereas a Mo filtration is needed to avoid the increases both in the low-energy component of the incident beam and in the higher energy component of the transmitted one. The contrast is improved with respect to the standard source and the mean glandular dose is only slightly increased for moderate transmission phantoms (thickness 2-4 cm).
引用
收藏
页码:1717 / 1726
页数:10
相关论文
共 13 条
[1]  
BARTOLOTTA A, 1988, PHYS MED, V3, P185
[2]   MAMMOGRAPHY WITH SYNCHROTRON-RADIATION [J].
BURATTINI, E ;
COSSU, E ;
DIMAGGIO, C ;
GAMBACCINI, M ;
INDOVINA, PL ;
MARZIANI, M ;
POCEK, M ;
SIMEONI, S ;
SIMONETTI, G .
RADIOLOGY, 1995, 195 (01) :239-244
[3]   X-RAY MAMMOGRAPHY WITH SYNCHROTRON RADIATION [J].
BURATTINI, E ;
GAMBACCINI, M ;
MARZIANI, M ;
RIMONDI, O ;
INDOVINA, PL ;
POCEK, M ;
SIMONETTI, G ;
BENASSI, M ;
TIRELLI, C ;
PASSARIELLO, R .
REVIEW OF SCIENTIFIC INSTRUMENTS, 1992, 63 (01) :638-640
[4]  
Calicchia A., 1994, Physica Medica, V10, P55
[5]  
CALICCHIA A, 1993, PHYS MEDICA, V9, P136
[6]   MONTE-CARLO CALCULATION OF CONVERSION FACTORS FOR THE ESTIMATION OF MEAN GLANDULAR BREAST DOSE [J].
DANCE, DR .
PHYSICS IN MEDICINE AND BIOLOGY, 1990, 35 (09) :1211-1219
[7]  
DANCE DR, 1981, P PATIENT EXPOSURE R, P227
[8]   INFLUENCE OF ANODE AND FILTER MATERIAL ON IMAGE QUALITY AND GLANDULAR DOSE FOR SCREEN FILM MAMMOGRAPHY [J].
DESPONDS, L ;
DEPEURSINGE, C ;
GRECESCU, M ;
HESSLER, C ;
SAMIRI, A ;
VALLEY, JF .
PHYSICS IN MEDICINE AND BIOLOGY, 1991, 36 (09) :1165-1182
[9]   NARROW ENERGY-BAND X-RAYS VIA MOSAIC CRYSTAL FOR MAMMOGRAPHY APPLICATION [J].
GAMBACCINI, M ;
TAIBI, A ;
DELGUERRA, A ;
FRONTERA, F ;
MARZIANI, M .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1995, 365 (01) :248-254
[10]  
*IPSM, 1989, COMM ROUT TEST MAMM