Methods for quantitative analysis of the primary stability in uncemented hip prostheses

被引:59
作者
Monti, L
Cristofolini, L
Viceconti, M
机构
[1] Univ Bologna, Fac Engn, I-40136 Bologna, Italy
[2] Ist Ortoped Rizzoli, Lab Tecnol Med, Bologna, Italy
关键词
experimental validation; uncemented hip prostheses; interface; micromotion measurement; mechanical testing; torsional stability;
D O I
10.1046/j.1525-1594.1999.06287.x
中图分类号
R318 [生物医学工程];
学科分类号
0831 [生物医学工程];
摘要
Torsional loads of daily activities contribute to the failure of the primary fixation of hip prostheses. Implant torsional stability must be evaluated prior to in vivo clinical trials. Whereas previous work has investigated this phenomenon, descriptions of physiologically accurate and reproducible in vitro methodologies are rare. The present study aimed to detect and control the sources of error and variability that influence in vitro methods. A typical set-up for the analysis of primary stability of hip stems was studied. Load cycles included proximal-to-distal axial force, torque, and bending moment. The effects of loading frequency and strain distribution across the cortical bone were investigated in order to optimize testing conditions and measurement set-up. The relative sheer motion at the bone-stem interface was measured transcortically using linear variable displacement transducers. The procedures developed for mounting specimens on the testing machine and for positioning sensors on the specimen were standardized and tested for reproducibility. The protocol was finally tested for repeatability and accuracy. Measurement errors were 2.3 mu m between load cycles and 4.9 mu m for repeated set-ups, comparing favorably with the literature.
引用
收藏
页码:851 / 859
页数:9
相关论文
共 41 条
[1]
[Anonymous], 1979, CLIN ORTHOP RELAT RE
[2]
IS STAIRCASE WALKING A RISK FOR THE FIXATION OF HIP IMPLANTS [J].
BERGMANN, G ;
GRAICHEN, F ;
ROHLMANN, A .
JOURNAL OF BIOMECHANICS, 1995, 28 (05) :535-553
[3]
BOSS JH, 1994, MED PROG TECHNOL, V20, P119
[4]
Histological and biomechanical analysis of bone and interface reactions around hydroxyapatite-coated intramedullary implants of different stiffness: a pilot study on the goat [J].
Buma, P ;
vanLoon, PJM ;
Versleyen, H ;
Weinans, H ;
Slooff, TJJH ;
deGroot, K ;
Huiskes, R .
BIOMATERIALS, 1997, 18 (18) :1251-1260
[5]
THE EFFECT OF FEMORAL STEM GEOMETRY ON INTERFACE MOTION IN UNCEMENTED POROUS-COATED TOTAL HIP PROSTHESES - COMPARISON OF STRAIGHT-STEM AND CURVED-STEM DESIGNS [J].
CALLAGHAN, JJ ;
FULGHUM, CS ;
GLISSON, RR ;
STRANNE, SK .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1992, 74A (06) :839-848
[6]
HYDROXYLAPATITE COATING OF POROUS IMPLANTS IMPROVES BONE INGROWTH AND INTERFACE ATTACHMENT STRENGTH [J].
COOK, SD ;
THOMAS, KA ;
DALTON, JE ;
VOLKMAN, TK ;
WHITECLOUD, TS ;
KAY, JF .
JOURNAL OF BIOMEDICAL MATERIALS RESEARCH, 1992, 26 (08) :989-1001
[7]
A critical analysis of stress shielding evaluation of hip prostheses [J].
Cristofolini, L .
CRITICAL REVIEWS IN BIOMEDICAL ENGINEERING, 1997, 25 (4-5) :409-483
[8]
Mechanical validation of whole bone composite femur models [J].
Cristofolini, L ;
Viceconti, M ;
Cappello, A ;
Toni, A .
JOURNAL OF BIOMECHANICS, 1996, 29 (04) :525-535
[9]
THE CEMENT MANTLE IN TOTAL HIP-ARTHROPLASTY - ANALYSIS OF LONG-TERM RADIOGRAPHIC RESULTS [J].
EBRAMZADEH, E ;
SARMIENTO, A ;
MCKELLOP, HA ;
LLINAS, A ;
GOGAN, W .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1994, 76A (01) :77-87
[10]
EBRAMZADEH E, 1988, J BONE JOINT SURG, V12, P464