摘要
The change in solution properties due to the agglutination of an antigen with its specific antibody has previously been used as a marker of infection. This method has been modified to allow the binding activity between species to be followed using the frequency response of a quartz crystal microbalance (QCM). The Bayston agglutination plate assay for Staphylococcus epidermidis has been modified to allow the electrode of a QCM to act as a direct sensor for the change in solution properties as agglutination occurs. Antibody and antigen were introduced to the crystal surface and the agglutination process was followed as a change in crystal resonant frequency. Serum, known to be infected with the organism, gave a titre of 3.9 x 10(-2)% v/v (-118 Hz, +/-12 SD, N = 9) matching that given by triplicate plate assay. Uninfected serum gave no frequency changes at this concentration, yielding a titre of 2.5 x 10(-2)% v/v again matching the plate titre (N = 3). Infected serum gave responses 40 times faster then those of the uninfected serum. The piezoelectric quartz crystal method gave a positive or negative diagnosis in < 15 min compared with the 24 h required for the plate assay. (C) 1999 Elsevier Science B.V. All rights reserved.
引用
收藏
页码:241 / 245
页数:5
相关论文
共 13 条
[1]   DEVELOPMENT OF QUARTZ CRYSTAL-OSCILLATORS FOR UNDER-LIQUID SENSING [J].
BARNES, C .
SENSORS AND ACTUATORS A-PHYSICAL, 1991, 29 (01) :59-69
[2]  
BAYSTON R, 1971, DEV MED CHILD NEUROL, V13, P135
[3]  
BLACK JA, 1965, LANCET, V2, P921
[4]  
Brock T.D., 1988, BIOL MICROORGANISMS
[5]   LATEX PIEZOELECTRIC IMMUNOASSAY - EFFECT OF INTERFACIAL PROPERTIES [J].
GHOURCHIAN, HO ;
KAMO, N .
ANALYTICA CHIMICA ACTA, 1995, 300 (1-3) :99-105
[7]  
KUROSAWA S, 1990, CHEM PHARM BULL, V38, P1117
[8]   QUARTZ-CRYSTAL GELATION DETECTOR FOR THE DETERMINATION OF FIBRINOGEN CONCENTRATION [J].
MURAMATSU, H ;
TAMIYA, E ;
SUZUKI, M ;
KARUBE, I .
ANALYTICA CHIMICA ACTA, 1989, 217 (02) :321-326
[9]   DETECTION OF ANTISTREPTOLYSIN O-ANTIBODY - APPLICATION OF AN INITIAL RATE METHOD OF LATEX PIEZOELECTRIC IMMUNOASSAY [J].
MURATSUGU, M ;
KUROSAWA, S ;
KAMO, N .
ANALYTICAL CHEMISTRY, 1992, 64 (21) :2483-2487
[10]   DEVELOPMENT OF A PIEZOELECTRIC IMMUNOSENSOR FOR THE DETECTION OF ENTEROBACTERIA [J].
PLOMER, M ;
GUILBAULT, GG ;
HOCK, B .
ENZYME AND MICROBIAL TECHNOLOGY, 1992, 14 (03) :230-235