Effect of different surface finishing and of hydroxyapatite coatings on passive and corrosion current of Ti6Al4V alloy in simulated physiological solution

被引:90
作者
Cabrini, M
Cigada, A
Rondelli, G
Vicentini, B
机构
[1] POLITECN MILAN,DIPARTIMENTO CHIM FIS APPL,I-20133 MILAN,ITALY
[2] UNIV AQUILA,DIPARTIMENTO CHIM INGN CHIM & MAT,I-67040 LAQUILA,ITALY
[3] CNR,IST TECNOL MAT,I-20133 MILAN,ITALY
关键词
passive and corrosion currents; Ti6Al4V; hydroxyapatite; plasma spray;
D O I
10.1016/S0142-9612(96)00205-0
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Direct and alternating current electrochemical tests were carried out on Ti6Al4V with different surface finishing and with hydroxyapatite (HA) coatings. Sand-blasting and rough titanium deposits obtained by vacuum plasma spraying (VPS) bring about an increase of passive and corrosion current density (c.d.) with respect to smooth Ti6Al4V, as a consequence of the augmentation of the real surface. The presence of HA deposits obtained by VPS causes an increase of passive and corrosion c.d. of the metallic substrate of about one order of magnitude and this should be taken into account in view of human body applications. (C) 1997 Elsevier Science Limited.
引用
收藏
页码:783 / 787
页数:5
相关论文
共 15 条
[1]  
BOUKAMP BA, 1989, CT88265128 U TWENT B
[2]   USE OF AC IMPEDANCE METHODS TO STUDY THE CORROSION BEHAVIOR OF IMPLANT ALLOYS [J].
BUNDY, KJ ;
DILLARD, J ;
LUEDEMANN, R .
BIOMATERIALS, 1993, 14 (07) :529-536
[3]  
BUNDY KJ, 1987, ASTM STP, V95, P137
[4]  
COOK SD, 1988, CLIN ORTHOP RELAT R, V230, P303
[5]  
De Levie R, 1964, ELECTROCHIM ACTA, V9, P1231, DOI DOI 10.1016/0013-4686(64)85015-5
[6]   THE EFFECT OF PLASMA-SPRAYED CALCIUM-PHOSPHATE CERAMIC COATINGS ON THE METAL-ION RELEASE FROM POROUS TITANIUM AND COBALT-CHROMIUM ALLOYS [J].
DUCHEYNE, P ;
HEALY, KE .
JOURNAL OF BIOMEDICAL MATERIALS RESEARCH, 1988, 22 (12) :1137-1163
[7]  
FRAKER AC, 1985, 853166 NBSIR
[8]  
GEESINK RGT, 1987, CLIN ORTHOP RELAT R, V225, P147
[9]   BIOCERAMICS - FROM CONCEPT TO CLINIC [J].
HENCH, LL .
JOURNAL OF THE AMERICAN CERAMIC SOCIETY, 1991, 74 (07) :1487-1510
[10]   CORROSION-RESISTANT ALLOYS IN CHLORIDE SOLUTIONS - MATERIALS FOR SURGICAL IMPLANTS [J].
HOAR, TP ;
MEARS, DC .
PROCEEDINGS OF THE ROYAL SOCIETY OF LONDON SERIES A-MATHEMATICAL AND PHYSICAL SCIENCES, 1966, 294 (1439) :486-&