Accuracy, precision and user-acceptability of self blood glucose monitoring machines

被引:37
作者
Chan, JCN
Wong, RYM
Cheung, CK
Lam, P
Chow, CC
Yeung, VTF
Kan, ECY
Loo, KM
Mong, MYL
Cockram, CS
机构
[1] CHINESE UNIV HONG KONG, PRINCE WALES HOSP, DIABET & ENDOCRINE CTR, SHATIN, NT, HONG KONG
[2] CHINESE UNIV HONG KONG, PRINCE WALES HOSP, DEPT CHEM PATHOL, SHATIN, HONG KONG
[3] CHINESE UNIV HONG KONG, PRINCE WALES HOSP, CTR CLIN TRIALS & EPIDEMIOL RES, SHATIN, HONG KONG
[4] CHINESE UNIV HONG KONG, PRINCE WALES HOSP, DEPT MED, SHATIN, HONG KONG
关键词
self-montoring; blood glucose; accuracy; precision; user-acceptability;
D O I
10.1016/S0168-8227(97)00036-3
中图分类号
R5 [内科学];
学科分类号
1002 [临床医学]; 100201 [内科学];
摘要
The performance of six self-monitoring blood glucose (SMBG) machines (Accutrend(R), Reflolux S(R), Companion 2(R), Glucometer GX(R), Glucometer IV(R) and One Touch II(R)) were examined using venous blood samples from 88 patients. Whole blood glucose (BG) values were measured by four machines from each brand. Machine-generated whole blood glucose (BG) values were corrected before comparison with laboratory plasma glucose values, measured by a glucose oxidase method. Based on error grid analysis, most of the corrected machine-generated BG values were clinically acceptable. Accutrend(R), Glucometer IV(R) and Companion 2(R) showed the greatest consistency between machines of the same brand. Over 80% of corrected BG values generated by Glucometer IV(R) fell within +/- 10% of the reference values. One Touch II(R) yielded the most reproducible results with a mean CV of 2.7% and was considered the most user friendly machine. More studies are required to examine the performance of these machines in the hands of patients. (C) 1997 Elsevier Science Ireland Ltd.
引用
收藏
页码:91 / 104
页数:14
相关论文
共 17 条
[1]
Anderson T. W., 1984, INTRO MULTIVARIATE S, P550
[2]
STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1986, 1 (8476) :307-310
[3]
CHAN AYW, 1989, CLIN CHEM, V35, P315
[4]
EVALUATING CLINICAL ACCURACY OF SYSTEMS FOR SELF-MONITORING OF BLOOD-GLUCOSE [J].
CLARKE, WL ;
COX, D ;
GONDERFREDERICK, LA ;
CARTER, W ;
POHL, SL .
DIABETES CARE, 1987, 10 (05) :622-628
[5]
*CONS PAN, 1996, DIABETES CARE, V19, pS62
[6]
DAVIDSON MB, 1981, DIABETES MELLITUS DI, V1, P814
[7]
DEVREESE K, 1993, EUR J CLIN CHEM CLIN, V31, P829
[8]
DUNN O, 1986, APPLIED STATISTICS A, P236
[9]
COMPARISON OF 5 GLUCOSE METERS FOR SELF-MONITORING OF BLOOD-GLUCOSE BY DIABETIC-PATIENTS [J].
GIFFORDJORGENSEN, RA ;
BORCHERT, J ;
HASSANEIN, R ;
TILZER, L ;
EAKS, GA ;
MOORE, WV .
DIABETES CARE, 1986, 9 (01) :70-76
[10]
GOLDBERG NJ, 1992, MEDISENSE