The implantable Cook-Swartz Doppler probe for postoperative monitoring in head and neck free flap reconstruction

被引:94
作者
Guillemaud, Jennifer P. [1 ]
Seikaly, Hadi [1 ]
Cote, David [1 ]
Allen, Heather [1 ]
Harris, Jeffrey R. [1 ]
机构
[1] Univ Alberta Hosp, Div Otolaryngol Head & Neck Surg, Walter MacKenzie Hlth Sci Ctr, Dept Surg, Edmonton, AB T6G 2R7, Canada
关键词
D O I
10.1001/archotol.134.7.729
中图分类号
R76 [耳鼻咽喉科学];
学科分类号
100213 [耳鼻咽喉科学];
摘要
Objective: To determine if the implantable Cook-Swartz Doppler Flow Monitoring System (Cook Vascular Inc, Vandergrift, Pennsylvania) improves surgical salvage rates for compromised free flaps. Design: Retrospective medical record review spanning 2002 to 2006 for a large head and neck oncology program. Setting: A tertiary care hospital. Patients: A consecutive series of 351 patients (244 men and 107 women; mean age, 58.63 years) who underwent free flap reconstruction of head and neck defects that were monitored using the implantable Doppler probe were included. Results: The most common indication for surgery was squamous cell carcinoma (81.0%), followed by functional reconstruction (4.3%). The most common free flap used was radial forearm (68.0%), followed by the fibular free flap (19.0%). With operative exploration used as the gold standard, the Cook-Swartz Doppler Flow Monitoring System had a sensitivity of 65.8% and specificity of 98.2% for the detection of flap compromise. For the detection of vascular compromise of the monitored vessel (excluding flap compromise cases whereby flow in the monitored vessel was not compromised on operative exploration, ie, venous obstruction, hematoma formation, and necrotizing fasciitis), the sensitivity increased to 100%. Conclusions: This is the largest reported series, to our knowledge, of implantable Cook-Swartz Doppler use, and our experience would suggest that this is a reliable technique for postoperative monitoring in head and neck reconstruction. Our use of the implantable Doppler probe allowed us to recognize vascular compromise early, resulting in an overall flap success rate of 98.1%, with a 92.0% salvage rate of flaps that experienced vascular compromise of the monitored vessel.
引用
收藏
页码:729 / 734
页数:6
相关论文
共 17 条
[1]
*AM JOINT COMM CAN, 2003, AJCC CANC STAG MAN
[2]
FREE FLAP MONITORING USING A MICROCOMPUTER LINKED TO A LASER DOPPLER FLOWMETER [J].
BRUCECHWATT, AJ .
BRITISH JOURNAL OF PLASTIC SURGERY, 1986, 39 (02) :229-238
[3]
de la Torre J, 2003, J RECONSTR MICROSURG, V19, P287
[4]
Disa JJ, 1999, PLAST RECONSTR SURG, V104, P97, DOI 10.1097/00006534-199907000-00014
[5]
MINIATURE IMPLANTABLE LASER DOPPLER PROBE MONITORING OF FREE TISSUE TRANSFER [J].
FERNANDO, B ;
YOUNG, VL ;
LOGAN, SE .
ANNALS OF PLASTIC SURGERY, 1988, 20 (05) :434-442
[6]
GRAHAM BH, 1983, PLAST RECONSTR SURG, V71, P826, DOI 10.1097/00006534-198306000-00016
[7]
TRANSACTION COST-ANALYSIS OF STRATEGY-STRUCTURE CHOICE [J].
JONES, GR ;
HILL, CWL .
STRATEGIC MANAGEMENT JOURNAL, 1988, 9 (02) :159-172
[8]
The effect of an implantable Doppler probe on the salvage of microvascular tissue transplants [J].
Kind, GM ;
Buntic, RF ;
Buncke, GM ;
Cooper, TM ;
Siko, PP ;
Buncke, HJ .
PLASTIC AND RECONSTRUCTIVE SURGERY, 1998, 101 (05) :1268-1273
[9]
Evaluation of 250 free-flap reconstructions after resection of tumours of the head and neck [J].
O'Brien, CJ ;
Lee, KK ;
Stern, HS ;
Traynor, SJ ;
Bron, L ;
Tew, PJ ;
Haghighi, KS .
AUSTRALIAN AND NEW ZEALAND JOURNAL OF SURGERY, 1998, 68 (10) :698-701
[10]
The Cook-Swartz venous doppler probe for the post-operative monitoring of free tissue transfers in the United Kingdom: a preliminary report [J].
Oliver, DW ;
Whitaker, IS ;
Giele, H ;
Critchley, P ;
Cassell, O .
BRITISH JOURNAL OF PLASTIC SURGERY, 2005, 58 (03) :366-370