Rapid measurement of the spatial resolution of colour flow scanners

被引:6
作者
Li, S [1 ]
Hoskins, PR [1 ]
McDicken, WN [1 ]
机构
[1] ROYAL INFIRM,DEPT MED PHYS & MED ENGN,EDINBURGH EH3 9YW,MIDLOTHIAN,SCOTLAND
关键词
colour velocity; Doppler ultrasound; spatial resolution;
D O I
10.1016/S0301-5629(97)00035-5
中图分类号
O42 [声学];
学科分类号
070206 [声学]; 082403 [水声工程];
摘要
An acoustic grid for rapid measurement of the spatial resolution of colour velocity images is described, The grid is composed of alternate widths of acoustic window and acoustic attenuator, Attenuation is provided by the use of a layer of waterproof paper, which is wrapped in close contact with a tube of a how phantom, The colour image consists of several closely spaced lines, When the grid line spacing is less than the spatial resolution, the colour lines cannot be separately visualised, The use of several grids of different spacing provides a measure of the spatial resolution, It is shown that the spatial resolution is dependent on the position of the focal zone, (C) 1997 World Federation for Ultrasound in Medicine & Biology.
引用
收藏
页码:591 / 596
页数:6
相关论文
共 11 条
[1]
FACE HEIGHT AND TOOTH ERUPTION IN ADULTS - A 20-YEAR FOLLOW-UP INVESTIGATION [J].
FORSBERG, CM ;
ELIASSON, S ;
WESTERGREN, H .
EUROPEAN JOURNAL OF ORTHODONTICS, 1991, 13 (04) :249-254
[2]
HALMSHAW R, 1981, PHYSICAL ASPECTS MED, P17
[3]
A SET OF X-RAY TEST OBJECTS FOR QUALITY-CONTROL IN TELEVISION FLUOROSCOPY [J].
HAY, GA ;
CLARKE, OF ;
COLEMAN, NJ ;
COWEN, AR .
BRITISH JOURNAL OF RADIOLOGY, 1985, 58 (688) :335-344
[4]
METHODS TO EVALUATE THE SAMPLE VOLUME OF PULSED DOPPLER SYSTEMS [J].
HOEKS, APG ;
RUISSEN, CJ ;
HICK, P ;
RENEMAN, RS .
ULTRASOUND IN MEDICINE AND BIOLOGY, 1984, 10 (04) :427-434
[5]
CHOICE OF MOVING TARGET FOR A STRING PHANTOM .2. ON THE PERFORMANCE TESTING OF DOPPLER ULTRASOUND SYSTEMS [J].
HOSKINS, PR .
ULTRASOUND IN MEDICINE AND BIOLOGY, 1994, 20 (08) :781-789
[6]
A COMPUTER-CONTROLLED FLOW PHANTOM FOR GENERATION OF PHYSIOLOGICAL DOPPLER WAVEFORMS [J].
HOSKINS, PR ;
ANDERSON, T ;
MCDICKEN, WN .
PHYSICS IN MEDICINE AND BIOLOGY, 1989, 34 (11) :1709-1717
[7]
CHOICE OF MOVING TARGET FOR A STRING PHANTOM .1. MEASUREMENT OF FILAMENT BACKSCATTER CHARACTERISTICS [J].
HOSKINS, PR .
ULTRASOUND IN MEDICINE AND BIOLOGY, 1994, 20 (08) :773-780
[8]
HOSKINS PR, 1994, TESTING DOPPLER ULTR
[10]
Price R., 1995, ROUTINE QUALITY ASSU