Eight-year results of hydroxyapatite-coated hip arthroplasty

被引:35
作者
Chang, Je-Ken
Chen, Chung-Hwan
Huang, Kuan-Yu
Wang, Gwo-Jaw
机构
[1] Kaohsiung Med Univ, Dept Orthoped, Kaohsiung 807, Taiwan
[2] Kaohsiung Med Univ, Dept Orthoped, Res Ctr, Kaohsiung 807, Taiwan
关键词
hydroxyapatite; total hip arthroplasty;
D O I
10.1016/j.arth.2005.04.043
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Ninety hips in 82 patients using Omnifit hydroxyapatite (HA)-coated prosthesis were followed for at least 7 years. All steins were stable at the final follow-up. However, aseptic loosening was found in 8 Cups and 6 of them were revised. Two polyethylene wear were treated with inserts exchanged. The mechanical failure rate was 11.4% and the combined failure rate was 14.3% for HA-coated cup. Four other cups with wear and osteolysis without loosening or pain and 2 cups with polyethylene wear without osteolysis were still under observation. Our findings suggest that hip arthroplasties with HA coating on the smooth surface of a titanium cup is not reliable. The mid-term result of HA-coated stern is as good as that of porous-coated stern.
引用
收藏
页码:541 / 546
页数:6
相关论文
共 25 条
[1]
High osteolysis and revision rate with the hydroxyapatite-coated ABG hip prostheses - 65 hips in 56 young patients followed for 5-9 years [J].
Blacha, J .
ACTA ORTHOPAEDICA SCANDINAVICA, 2004, 75 (03) :276-282
[2]
Clinical and radiological results of hydroxyapatite-coated acetabular cups [J].
Blacha, J ;
Gagala, J .
INTERNATIONAL ORTHOPAEDICS, 2004, 28 (06) :362-365
[3]
CAPELLO WN, 1994, ORTHOPEDICS, V17, P781
[4]
Hydroxyapatite in total hip arthroplasty -: Clinical results and critical issues [J].
Capello, WN ;
D 'Antonio, JA ;
Manley, MT ;
Feinberg, JR .
CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 1998, (355) :200-211
[5]
Bone ingrowth on a smooth-surfaced hydroxyapatite-coated acetabular cup [J].
Chung, YY ;
Kim, HD ;
Kim, KS .
INTERNATIONAL ORTHOPAEDICS, 2002, 26 (05) :283-286
[6]
D'Antonio JA, 2001, CLIN ORTHOP RELAT R, P101
[7]
DAntonio JA, 1997, CLIN ORTHOP RELAT R, P124
[8]
EARLY CLINICAL-EXPERIENCE WITH HYDROXYAPATITE-COATED FEMORAL IMPLANTS [J].
DANTONIO, JA ;
CAPELLO, WN ;
CROTHERS, OD ;
JAFFE, WL ;
MANLEY, MT .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1992, 74A (07) :995-1008
[9]
DELEE JG, 1976, CLIN ORTHOP RELAT R, P20
[10]
ENGH CA, 1990, CLIN ORTHOP RELAT R, P107