ECG measurement on a chair without conductive contact

被引:171
作者
Lim, YG [1 ]
Kim, KK [1 ]
Park, KS [1 ]
机构
[1] Seoul Natl Univ, Coll Med, Dept Biomed Engn, Seoul 100799, South Korea
关键词
active electrode; ECG monitoring; high-input impedance; indirect-contact electrode;
D O I
10.1109/TBME.2006.872823
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
For the purpose of long-term, everyday electrocardiogram (ECG) monitoring, we present a convenient method of ECG measurement without direct conductive contact with the skin while subjects sat on a chair wearing normal clothes. Measurements were made using electrodes attached to the back of a chair, high-input-impedance amplifiers mounted on the electrodes, and a large ground-plane placed on the chair seat. ECGs were obtained by the presented method for several types of clothing and compared to ECGs obtained from conventional measurement using Ag-AgCl electrodes. Motion artifacts caused by usual desk works were investigated. This study shows the feasibility of the method for long-term, convenient, everyday use.
引用
收藏
页码:956 / 959
页数:4
相关论文
共 6 条
[1]  
*ANSI STAND, EC111991 ANSIAAMI
[2]   Electric potential probes - new directions in the remote sensing of the human body [J].
Harland, CJ ;
Clark, TD ;
Prance, RJ .
MEASUREMENT SCIENCE AND TECHNOLOGY, 2002, 13 (02) :163-169
[3]   MONITORING OF ELECTROCARDIOGRAMS IN BED WITHOUT UTILIZING BODY-SURFACE ELECTRODES [J].
ISHIJIMA, M .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1993, 40 (06) :593-594
[4]   Attempts at monitoring health status in the home [J].
Ogawa, M ;
Togawa, T .
1ST ANNUAL INTERNATIONAL IEEE-EMBS SPECIAL TOPIC CONFERENCE ON MICROTECHNOLOGIES IN MEDICINE & BIOLOGY, PROCEEDINGS, 2000, :552-556
[5]  
OTT HW, 1988, NOISE REDUCTION TECH, P244
[6]   Fully automated health monitoring system in the home [J].
Tamura, T ;
Togawa, T ;
Ogawa, M ;
Yoda, R .
MEDICAL ENGINEERING & PHYSICS, 1998, 20 (08) :573-579