Biomechanical comparison of the sandblasted and acid-etched and the machined and acid-etched titanium surface for dental implants

被引:205
作者
Li, DH
Ferguson, SJ
Beutler, T
Cochran, DL
Sittig, C
Hirt, HP
Buser, D
机构
[1] Univ Bern, Sch Dent Med, Dept Oral Surg & Stomatol, CH-3010 Bern, Switzerland
[2] Univ Bern, ME Muller Inst Biomech, CH-3010 Bern, Switzerland
[3] Inst Straumann AG, CH-4437 Waldenburg, Switzerland
[4] Univ Texas, Hlth Sci Ctr, Dept Periodont, San Antonio, TX USA
来源
JOURNAL OF BIOMEDICAL MATERIALS RESEARCH | 2002年 / 60卷 / 02期
关键词
removal torque values; interfacial shear strength; interfacial stiffness; sandblasted and acid-etched surface; acid-etched surface; titanium implants;
D O I
10.1002/jbm.10063
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
To make a direct biomechanical comparison between the sandblasted and acid-etched surface (SLA) and the machined and acid-etched surface (MA), a well-established animal model for implant removal torque testing was employed, using a split-mouth experimental design. All implants had an identical cylindrical solid-screw shape with the standard ITI thread configuration, without any macroscopic retentive structures. After 4, 8, and 12 weeks of bone healing, removal torque testing was performed to evaluate the interfacial shear strength of each surface type. Results showed that the SLA surface was more powerful in enhancing the interfacial shear strength of implants in comparison with the MA surface. Removal torque values of the SLA-surfaced implants were about 30% higher than those of the MA-surfaced implants (p = 0.002) except at 4 weeks, when the difference was at the threshold of statistical significance (p = 0.0519). The mean removal torque values for the SLA implants were 1.5074 Nm at 4 weeks, 1.8022 Nm at 8 weeks, and 1.7130 Nm at 12 weeks; and correspondingly, 1.1924 Nm, 1.3092 Nm, and 1.3226 Nm for the MA implants. It can be concluded that the SLA surface achieves a better bone anchorage than the MA surface, and that sandblasting before acid etching has a beneficial effect on the interfacial shear strength. As regards the bone-implant interfacial stiffness calculated from the torque-rotation curve, the SLA implants showed an overall more than 5% higher stiffness compared with the MA implants, although the difference did not reach the statistical significance level. (C) 2002 John Wiley Sons, Inc.
引用
收藏
页码:325 / 332
页数:8
相关论文
共 39 条
[1]   A 15-YEAR STUDY OF OSSEOINTEGRATED IMPLANTS IN THE TREATMENT OF THE EDENTULOUS JAW [J].
ADELL, R ;
LEKHOLM, U ;
ROCKLER, B ;
BRANEMARK, PI .
INTERNATIONAL JOURNAL OF ORAL SURGERY, 1981, 10 (06) :387-416
[2]  
ADELL R, 1990, International Journal of Oral and Maxillofacial Implants, V5, P347
[3]   TITANIUM PLASMA-SPRAYED (TPS) SCREW IMPLANTS FOR THE RECONSTRUCTION OF THE EDENTULOUS MANDIBLE [J].
BABBUSH, CA ;
KENT, JN ;
MISIEK, DJ .
JOURNAL OF ORAL AND MAXILLOFACIAL SURGERY, 1986, 44 (04) :274-282
[4]  
Baker D, 1999, INT J ORAL MAX IMPL, V14, P722
[5]  
Boyan BD, 1998, J BIOMED MATER RES, V39, P77, DOI 10.1002/(SICI)1097-4636(199801)39:1<77::AID-JBM10>3.0.CO
[6]  
2-L
[7]  
Branemark P I, 1977, Scand J Plast Reconstr Surg Suppl, V16, P1
[8]  
BRANEMARK P I, 1969, Scandinavian Journal of Plastic and Reconstructive Surgery, V3, P81
[9]  
Buser D, 1998, INT J ORAL MAX IMPL, V13, P611
[10]   Tissue integration of non-submerged implants 1-year results of a prospective study with 100 ITI hollow-cylinder and hollow-screw implants [J].
Buser, D. ;
Weber, H-P ;
Lang, N. P. .
CLINICAL ORAL IMPLANTS RESEARCH, 1990, 1 (01) :33-40