Active optical thin-film waveguide sensor for ion sensing

被引:16
作者
Kim, KH
Minamitani, H
Hisamoto, H
Suzuki, K
Kang, SW
机构
[1] KEIO UNIV,FAC SCI & TECHNOL,DEPT APPL CHEM,YOKOHAMA,KANAGAWA 223,JAPAN
[2] KYUNGPOOK NATL UNIV,STRC,TAEGU 702701,SOUTH KOREA
关键词
sensors; optical ion sensor; optode; waveguide;
D O I
10.1016/S0003-2670(97)00006-8
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
In this paper, an active PVC optical thin-film waveguide sensor is proposed which can select a specific ion and measure its concentration. The waveguide was fabricated using the plasticized poly(vinyl chloride) (PVC), bis ethyl hexyl phthalate (BEHP), lipophilic anionic dye (LAD-3), and ionophores (K23E1 for calcium, or C14DD16C5 for sodium) in which the PVC membrane plays important roles for both incident light propagation and selective ion sensing. In order to investigate the applicability of this sensor to Ca2+ and Na+ measurement, the sensor was applied to sample solutions whose CaCl2 concentration was gradually altered. Also, we applied it to a NaCl solution. In this experiment, we found that this PVC thin-film waveguide possessed good light propagation for argon laser light at 488 nm. Also, the comparison of the experimental responses between the active waveguide sensor and a commonly utilized waveguide sensor (evanescent waveguide spectrometry), indicated that the sensitivity of the active waveguide sensor is obviously higher than that of the conventional waveguide sensor. These findings suggested that the proposed PVC waveguide ion sensor is applicable to the determination of various ions with high sensitivity.
引用
收藏
页码:199 / 208
页数:10
相关论文
共 28 条
[1]   SELECTIVITY OF ION-SENSITIVE BULK OPTODES [J].
BAKKER, E ;
SIMON, W .
ANALYTICAL CHEMISTRY, 1992, 64 (17) :1805-1812
[2]   PLANAR WAVE-GUIDE IMMUNOSENSOR WITH FLUORESCENT LIPOSOME AMPLIFICATION [J].
CHOQUETTE, SJ ;
LOCASCIOBROWN, L ;
DURST, RA .
ANALYTICAL CHEMISTRY, 1992, 64 (01) :55-60
[3]  
CHRISTENSEN D, 1993, P SOC PHOTO-OPT INS, V1886, P2, DOI 10.1117/12.144833
[4]   THIN-FILM PLANAR WAVE-GUIDE SENSOR FOR LIQUID-PHASE ABSORBENCY MEASUREMENTS [J].
DEGRANDPRE, MD ;
BURGESS, LW ;
WHITE, PL ;
GOLDMAN, DS .
ANALYTICAL CHEMISTRY, 1990, 62 (18) :2012-2017
[5]  
DOHNER RE, 1992, CHIMIA, V46, P215
[6]  
Harrick N.J., 1979, INTERNAL REFLECTION
[7]   NOVEL TYPE OF ION-SELECTIVE FLUOROSENSOR BASED ON THE INNER FILTER EFFECT - AN OPTRODE FOR POTASSIUM [J].
HE, HR ;
LI, H ;
MOHR, G ;
KOVACS, B ;
WERNER, T ;
WOLFBEIS, OS .
ANALYTICAL CHEMISTRY, 1993, 65 (02) :123-127
[8]  
HERRON JN, 1993, P SOC PHOTO-OPT INS, V1885, P28, DOI 10.1117/12.144735
[9]  
HISAMOTO H, 1995, SENSOR ACTUAT B-CHEM, V29, P378, DOI 10.1016/0925-4005(95)01711-9
[10]   FLOW-THROUGH TYPE CALCIUM-ION SELECTIVE OPTODES BASED ON NOVEL NEUTRAL IONOPHORES AND A LIPOPHILIC ANIONIC DYE [J].
HISAMOTO, H ;
WATANABE, K ;
NAKAGAWA, E ;
SISWANTA, D ;
SHICHI, Y ;
SUZUKI, K .
ANALYTICA CHIMICA ACTA, 1994, 299 (02) :179-187