Warning on inaccurate Rosenbaum cards for testing near vision

被引:30
作者
Horton, JC
Jones, MR
机构
[1] Beckman Vision Center, University of California, San Francisco, CA
[2] Beckman Vision Center, Univ. of California, San Francisco, San Francisco, CA 94143-0730
关键词
Jaeger; Snellen; near vision; Rosenbaum card; visual acuity;
D O I
10.1016/S0039-6257(97)00055-6
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
The Rosenbaum card is the most widely used handheld card for measuring near visual acuity. It was developed by Dr. J. George Rosenbaum of Cleveland, Ohio, for testing vision at the bedside of patients after-cataract surgery. Millions of copies of the Rosenbaum card have been distributed free by drug companies or sold by medical supply firms. Most versions of the Rosenbaum card are inaccurate because the numbers are not scaled properly to the Snellen system. This article reviews the history of the Rosenbaum card, briefly summarizes issues raised by near vision testing, and provides standards for the manufacture of Rosenbaum near cards. (C) 1997 by Elsevier Science Inc. All rights reserved.