A high efficiency device for the detection of radioactive xenon isotopes

被引:9
作者
Lamaze, GP
机构
[1] CSTL, NIST, Gaithersburg
关键词
D O I
10.1016/S0168-9002(96)01149-7
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
A detector has been designed for the detection of sub-Becquerel amounts of the radioactive xenon isotopes Xe-131m and Xe-133. High efficiency has been obtained by the condensation of the sample directly onto the surfaces of two closely spaced silicon surface-barrier detectors. The measurement of the Xe-131m is based onthe detection of the conversion electrons. The Xe-133 measurement is obtained by detection of the 81 keV photon in a NaI detector that nearly surrounds the silicon detectors. The minimum detectable activity for Xe-133 is 0.015 Bq. For a Xe-113m sample with no Xe-133 present, the minimum detectable activity is about 0.003 Bq.
引用
收藏
页码:285 / 289
页数:5
相关论文
共 6 条
[1]   A SYSTEM FOR RAPID CHEMICAL SEPARATION AND SPECTROSCOPIC EXAMINATION OF FISSION GASES [J].
ALEXANDER, P ;
LAU, JP ;
SAHAGIAN, AV .
NUCLEAR INSTRUMENTS & METHODS, 1968, 64 (03) :323-+
[2]   A 4PIBETA-SPECTROMETER WITH LI-SI COUNTERS [J].
ANDERSEN, V ;
CHRISTENSEN, CJ .
NUCLEAR INSTRUMENTS & METHODS, 1968, 61 (01) :77-+
[3]   4PI BETA-GAMMA-COINCIDENCE SPECTROMETER USING SI(LI) AND NAI(TL) DETECTORS [J].
GILS, HJ ;
WIESNER, W ;
LOHKEN, R ;
FLOTHMAN.D .
NUCLEAR INSTRUMENTS & METHODS, 1972, 105 (02) :179-&
[4]  
KNOP G, 1965, ALPHA BETA GAMMA RAY, P11
[5]  
SERGEENKOV YV, 1986, NUCL DATA SHEETS, V49, P689
[6]  
SERGEENKOV YV, 1994, NUCL DATA SHEETS, V72, P537