A three-dimensional volumetric test object for geometry evaluation in magnetic resonance imaging

被引:18
作者
McRobbie, DW
机构
[1] Radiological Sciences Unit, Hammersmith Hospitals NHS Trust, Charing Cross Hospital, London W6 8RF, Fulham Palace Road
关键词
magnetic resonance imaging; image quality; test objects; three-dimensional imaging;
D O I
10.1118/1.598123
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The design of a test object for the volumetric evaluation of geometric image quality parameters in magnetic resonance imaging (MRI) is presented. Two-dimensional parameters: linearity, distortion, slice position, slice width, slice oblique angle, spatial resolution, line spread function (LSF) are measurable in all slice planes within a volume corresponding to a typical head examination. Additionally, parameters particular to three-dimensional Fourier transform (3-DFT) imaging are measurable: 3D ghosting and 3D LSF. Parameter measurements are correctable with regard to test object positioning through the determination of the six degrees of freedom of the test object within the scanner's field of view. (C) 1997 American Association of Physicists in Medicine.
引用
收藏
页码:737 / 742