A simple model for an amperometric channel biosensor

被引:13
作者
Gooding, JJ [1 ]
机构
[1] Univ New S Wales, Sch Chem, Sydney, NSW 2052, Australia
关键词
biosensors; channel flow cells; enzyme electrodes;
D O I
10.1016/S1388-2481(99)00015-6
中图分类号
O646 [电化学、电解、磁化学];
学科分类号
081704 ;
摘要
A simple model is presented for the channel biosensor where an oxidase enzyme layer is located upstream of a detector electrode. In this model the enzyme kinetics are restricted to the linear region. The model enables the elucidation of the parameters important in tuning the response of the biosensor. Two extreme regimes of operation are identified; the kinetically limited and mass transport limited regimes both provide features which overcome reproducibility problems associated with the classical enzyme electrode geometry. (C) 1999 Elsevier Science S.A. All rights reserved.
引用
收藏
页码:119 / 123
页数:5
相关论文