Clinical comparison of the Treponema pallidum CAPTIA syphilis-G enzyme immunoassay with the fluorescent treponemal antibody absorption immunoglobulin G assay for syphilis testing
被引:22
作者:
Halling, VW
[1
]
Jones, MF
[1
]
Bestrom, JE
[1
]
Wold, AD
[1
]
Rosenblatt, JE
[1
]
Smith, TF
[1
]
Cockerill, FR
[1
]
机构:
[1] Mayo Clin, Div Clin Microbiol, Rochester, MN 55905 USA
关键词:
D O I:
10.1128/JCM.37.10.3233-3234.1999
中图分类号:
Q93 [微生物学];
学科分类号:
071005 ;
100705 ;
摘要:
Recently, a treponema-specific immunoglobulin G (IgG) enzyme immunoassay (EIA), the CAPTIA Syphilis-G (Trinity Biotech, Jamestown, N.Y.), has become available as a diagnostic test for syphilis. A total of 89 stored sera previously tested by the fluorescent treponemal antibody absorption (FTA-ABS) IgG assay were evaluated by the CAPTIA EIA, The PTA-ABS IgG procedure was performed by technologists unblinded to results of rapid plasmid reagin (RPR) testing of the same specimens. Borderline CAPTIA-positive samples (antibody indices of greater than or equal to 0.650 and less than or equal to 0.900) were retested; if the second analysis produced an index of > 0.900, the sample was considered positive, Thirteen of 89 (15%) samples had discrepant results. Compared to the FTA-ABS assay, the CAPTIA EIA had a sensitivity and specificity and positive and negative predictive values of 70.7, 97.9, 96.7, and 79.7%, respectively. In another analysis, discrepancies between results were resolved by repeated FTA-ABS testing (technologists were blinded to previous RPR results) and patient chart reviews. Seven CAPTIA-negative samples which were previously interpreted (unblinded) as minimally reactive by the PTA method were subsequently interpreted (blinded) as nonreactive, One other discrepant sample (CAPTIA negative and FTA-ABS positive [at an intensity of 3+], unblinded) was FTA negative with repeated testing (blinded), For the five remaining discrepant samples, chart reviews indicated that one patient (CAPTIA negative and FTA-ABS positive [minimally reactive], blinded) had possible syphilis. These five samples were also evaluated and found to be negative by another treponema-specific test, the Treponema pallidum microhemagglutination assay, Therefore, after repeated testing and chart reviews, 2 of the 89 (2%) samples had discrepant results; the adjusted sensitivity, specificity, and positive and negative predictive values were 96.7, 98.3, 96.7, and 98.3%, respectively. This study demonstrates that the CAPTIA Ige EIA is a reliable method for syphilis testing and that personnel performing tests which require subjective interpretation, like the FTA-ABS test, may be biased by RPR test results.
引用
收藏
页码:3233 / 3234
页数:2
相关论文