Computer aided speckle interferometry: A technique for measuring deformation of the surface of the heart

被引:28
作者
Gaudette, GR
Todaro, J
Krukenkamp, IB
Chiang, FP [1 ]
机构
[1] SUNY Stony Brook, Dept Mech Engn, Stony Brook, NY 11794 USA
[2] SUNY Stony Brook, Div Cardiothorac Surg, Stony Brook, NY 11794 USA
[3] SUNY Stony Brook, Univ Hosp Stony Brook, Stony Brook, NY 11794 USA
[4] SUNY Stony Brook, Inst Mol Cardiol, Stony Brook, NY 11794 USA
基金
美国国家科学基金会;
关键词
myocardial mechanics; speckle interferometry; epicardial deformation; sonomicrometry; soft tissue mechanics;
D O I
10.1114/1.1397785
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
An investigation of the inhomogeneous and anisotropic properties of myocardium necessitates a whole field measurement technique with high spatial resolution. Computer aided speckle interferometry (CASI) may be applied to measuring deformation on the epicardial surface of the heart. Silicone carbide particles (approximately 40 mum in diameter) were sprinkled randomly onto the epicardial surface of isolated rabbit hearts. When illuminated with white light, speckles may be observed with a charge coupled device (CCD) camera. A balloon was placed in the left ventricle to control the intracavitary load on the arrested heart. To compare CASI to the "gold" standard technique of sonomicrometry, two ultrasonic transducers were implanted into the wall of the myocardium. Three hearts were exposed to various loading conditions, and at each condition speckle images were recorded. CASI was used to determine the distribution of displacement vectors (both direction and magnitude) in the region imaged by the CCD camera. Strain along the axis of the implanted transducers was determined with CAST and compared to that obtained with sonomicrometry. Strain determined from CAST and sonomicrometry produced equivalent results. Unlike sonomicrometry, whereby the displacement between two points with a relatively large gauge length is obtained, CASI is able to determine displacement vectors for hundreds of "points" within the same region. In conclusion, CASI produced equivalent results to those obtained from sonomicrometry (although not with the same temporal resolution), but it is a whole field deformation mapping technique that has a spatial resolution three orders of magnitude higher than that of sonomicrometry. (C) 2001 Biomedical Engineering Society.
引用
收藏
页码:775 / 780
页数:6
相关论文
共 29 条
[1]   THEORY AND APPLICATIONS OF THE WHITE-LIGHT SPECKLE METHOD FOR STRAIN ANALYSIS [J].
ASUNDI, A ;
CHIANG, FP .
OPTICAL ENGINEERING, 1982, 21 (04) :570-580
[2]   DOES CARDIOPULMONARY BYPASS ALONE ELICIT MYOPROTECTIVE PRECONDITIONING [J].
BURNS, PG ;
KRUKENKAMP, IB ;
CALDARONE, CA ;
GAUDETTE, GR ;
BUKHARI, EA ;
LEVITSKY, S .
CIRCULATION, 1995, 92 (09) :447-451
[3]   DIGITAL SPECKLE-DISPLACEMENT MEASUREMENT USING A COMPLEX SPECTRUM METHOD [J].
CHEN, DJ ;
CHIANG, FP ;
TAN, YS ;
DON, HS .
APPLIED OPTICS, 1993, 32 (11) :1839-1849
[4]  
CHEN DJ, 1991, P 2 INT C PHOT SPECK, pA1554
[5]  
CHEN DJ, 1993, THESIS SUNY STONY BR
[6]  
CHEN DJ, 1989, P SEM ANN SPR C EXP
[7]  
CHIANG FP, 1978, SOLID MECH ARCH, V30, P1
[8]  
CHIANG FP, 1989, METALS HDB, V17, P432
[9]  
CHIANG FP, 1997, ASTM, P156
[10]  
Erf R.K., 1978, SPECKLE METROLOGY