Evaluation of commercially available fluid-warming devices for use in forward surgical and combat areas

被引:16
作者
Dubick, MA [1 ]
Brooks, DE [1 ]
Macaitis, JM [1 ]
Bice, TG [1 ]
Moreau, AR [1 ]
Holcomb, JB [1 ]
机构
[1] USA, Inst Surg Res, Ft Sam Houston, TX 78234 USA
关键词
D O I
10.7205/MILMED.170.1.76
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
The fluid-warming capabilities of four individual fluid warmers, i.e., Level 1, FMS 2000, Thermal Angel, and Ranger, were compared to evaluate their potential for medical use in forward military echelons of care. Lactated Ringer's solution (LR) and Hextend at room temperature (20 degrees C) or refrigerated temperature (4-7 degrees C) and packed red blood cells at 4 degrees C to 7 degrees C were used with each warmer at two different flow rates. The FMS 2000 consistently warmed all fluids to approximately 37 degrees C, regardless of the starting temperature or flow rate. The Level 1 and Ranger also efficiently warmed all fluids except cold LR to approximately 37 degrees C. The Thermal Angel generally warmed room temperature fluid, cold Hextend, and packed red blood cells to at least 33 degrees C to 34 degrees C but could not warm cold LR. The clinical standard is to have fluids warmed to 32 degrees C at a minimum and more preferably to 34 degrees C to 35 degrees C. Of the fluid warmers tested, only the Thermal Angel failed to achieve such a temperature in warming cold LR. Data from the present study suggest the Ranger and FMS 2000 to be operationally adaptable to at least echelons 1 and 2, respectively, whereas far-forward use of the Thermal Angel has limitations.
引用
收藏
页码:76 / 82
页数:7
相关论文
共 22 条
[1]   AN EVALUATION OF THE LEVEL-1 BLOOD WARMER SERIES [J].
BROWNE, DA ;
DEBOECK, R ;
MORGAN, M .
ANAESTHESIA, 1990, 45 (11) :960-963
[2]   COMPARISON OF 2 IN-LINE BLOOD WARMERS [J].
CHERRY, MS ;
HODGSON, GH ;
NOTTEBROCK, H .
CANADIAN ANAESTHETISTS SOCIETY JOURNAL, 1981, 28 (02) :180-181
[3]  
COMUNALE ME, 2000, TODAYS THER TRENDS, V18, P87
[4]   Hypothermia, coagulopathy, and acidosis [J].
Eddy, VA ;
Morris, JA ;
Cullinane, DC .
SURGICAL CLINICS OF NORTH AMERICA, 2000, 80 (03) :845-+
[5]   TEMPERATURE RELATIONSHIP TO DISTANCE AND FLOW-RATE OF WARMED-IV FLUIDS [J].
FARIES, G ;
JOHNSTON, C ;
PRUITT, KM ;
PLOUFF, RT .
ANNALS OF EMERGENCY MEDICINE, 1991, 20 (11) :1198-1200
[6]   Effects of short heat exposure on human red and white blood cells [J].
Fildes, J ;
Fisher, S ;
Sheaff, CM ;
Barrett, JA .
JOURNAL OF TRAUMA-INJURY INFECTION AND CRITICAL CARE, 1998, 45 (03) :479-483
[7]   ADVANCES IN THE MANAGEMENT OF HYPOTHERMIA [J].
GENTILELLO, LM .
SURGICAL CLINICS OF NORTH AMERICA, 1995, 75 (02) :243-256
[8]   Is hypothermia in the victim of major trauma protective or harmful? A randomized, prospective study [J].
Gentilello, LM ;
Jurkovich, GJ ;
Stark, MS ;
Hassantash, SA ;
OKeefe, GE .
ANNALS OF SURGERY, 1997, 226 (04) :439-447
[9]   Factors and methodology in achieving ideal delivery temperatures for intravenous and lavage fluid in hypothermia [J].
Handrigan, MT ;
Wright, RO ;
Becker, BM ;
Linakis, JG ;
Jay, GD .
AMERICAN JOURNAL OF EMERGENCY MEDICINE, 1997, 15 (04) :350-353
[10]   The limits of bloodwarming: Maximally heating blood with an inline microwave bloodwarmer [J].
Herron, DM ;
Grabowy, R ;
Connolly, R ;
Schwaitzberg, SD .
JOURNAL OF TRAUMA-INJURY INFECTION AND CRITICAL CARE, 1997, 43 (02) :219-226