Comparison of nonmydriatic digitized video fundus images with standard 35-mm slides to screen for and identify specific lesions of age-related macular degeneration

被引:11
作者
Lim, JI
Labree, L
Nichols, T
Cardenas, I
机构
[1] Doheny Eye Inst, Los Angeles, CA USA
[2] Univ So Calif, Dept Prevent Med Stat Consultat, Los Angeles, CA 90089 USA
[3] Univ So Calif, Res Ctr, Los Angeles, CA 90089 USA
来源
RETINA-THE JOURNAL OF RETINAL AND VITREOUS DISEASES | 2002年 / 22卷 / 01期
关键词
digitized video fundus images; 35-mm slides; age-related macular degeneration;
D O I
10.1097/00006982-200202000-00011
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Purpose: The purpose of this study was to compare nonmydriatic digitized images obtained using a digital imaging system (resolution of 640 X 480 pixels) with 35-mm slide images for detecting specific findings of age-related macular degeneration (AMD) and to evaluate its usefulness as a screening tool in detecting signs of AMD. Methods: Seventeen consecutive patients (33 eyes) underwent digital color imaging (with a nonmydriatic, 45-degree, fundus camera attached to a digital back) and standard 35-mm, 30-degree retinal color photography of the fundus: posterior pole, nasal retina, and temporal retina. The images were later reviewed for the presence or absence of specific retinal findings. The images were not compressed. Primary outcome measures included the presence or absence of drusen, hard exudate, choroidal neovascularization (CNV), subretinal hemorrhage, retinal pigment epithelial (RPE) changes, subretinal fibrosis, Pigment epithelial detachment (PED), and subretinal fluid. Presence of drusen, with or without any one of the other findings, and presence of disciform scar or geographic atrophy were positive indications in screening for AMD. Results: Agreement between image type was highest for FED (97%), CNV (91 0/o), and subretinal fibrosis (91%); and lowest for RPE changes (63%). Sensitivity, specificity, positive predictive value, and negative predictive values were determined using the 35 mm-slide images as the reference for comparison. Sensitivity ranged from 40% (hard exudates) to 75% (subretinal hemorrhage). Specificity ranged from 88% (drusen) to 100% (hard exudate, CNV, RPE changes, PED). Positive predictive value ranged from 67% (subretinal hemorrhage) to 100% (hard exudate, CNV, RPE changes, PED). Negative predictive value ranged from 44% (drusen) to 97% (PED). For the purposes of screening for any evidence of AMD, the system was 70% sensitive. Conclusion: This digital fundus imaging system with 640 X 480 pixel resolution has low sensitivity and high specificity, as compared with 35-mm slide images, for detection of early AMD, but higher sensitivity for late findings (CNV, scar, atrophy) of AMD. Because of sensitivity for detecting any AMD coupled with the low sensitivity for detecting CNV, the system is not useful for evaluating AMD patients who require close follow-up and who are at risk for more severe visual loss.
引用
收藏
页码:59 / 64
页数:6
相关论文
共 10 条
[1]   AN INTERNATIONAL CLASSIFICATION AND GRADING SYSTEM FOR AGE-RELATED MACULOPATHY AND AGE-RELATED MACULAR DEGENERATION [J].
BIRD, AEC ;
BRESSLER, NM ;
BRESSLER, SB ;
CHISHOLM, IH ;
COSCAS, G ;
DAVIS, MD ;
DEJONG, PTVM ;
KLAVER, CCW ;
KLEIN, BEK ;
KLEIN, R ;
MITCHELL, P ;
SARKS, JP ;
SARKS, SH ;
SOURBANE, G ;
TAYLOR, HR ;
VINGERLING, JR .
SURVEY OF OPHTHALMOLOGY, 1995, 39 (05) :367-374
[2]   Stereo nonmydriatic digital-video color retinal imaging compared with early treatment diabetic retinopathy study seven standard field 35-mm stereo color photos for determining level of diabetic retinopathy [J].
Bursell, SE ;
Cavallerano, JD ;
Cavallerano, AA ;
Clermont, AC ;
Birkmire-Peters, D ;
Aiello, LP ;
Aiello, LM .
OPHTHALMOLOGY, 2001, 108 (03) :572-585
[3]  
Flowers C W Jr, 1997, Telemed J, V3, P43, DOI 10.1089/tmj.1.1997.3.43
[4]   THE QUALITY OF PHOTOGRAPHS PRODUCED BY THE NONMYDRIATIC FUNDUS CAMERA IN A SCREENING-PROGRAM FOR DIABETIC-RETINOPATHY - A 1 YEAR PROSPECTIVE-STUDY [J].
HEAVEN, CJ ;
CANSFIELD, J ;
SHAW, KM .
EYE, 1993, 7 :787-790
[5]  
KLEIN BEK, 1992, OPHTHALMOLOGY, V99, P546
[6]  
KLEIN R, 1992, OPHTHALMOLOGY, V99, P1686
[7]   MEASUREMENT OF OBSERVER AGREEMENT FOR CATEGORICAL DATA [J].
LANDIS, JR ;
KOCH, GG .
BIOMETRICS, 1977, 33 (01) :159-174
[8]   Telemedicine and ophthalmology [J].
Li, HK .
SURVEY OF OPHTHALMOLOGY, 1999, 44 (01) :61-72
[9]   A comparison of digital nonmydriatic fundus imaging with standard 35-millimeter slides for diabetic retinopathy [J].
Lim, JI ;
LaBree, L ;
Nichols, T ;
Cardenas, I .
OPHTHALMOLOGY, 2000, 107 (05) :866-870
[10]   Telemedicine diagnosis of eye disorders by direct ophthalmoscopy - A pilot study [J].
Marcus, DM ;
Brooks, SE ;
Ulrich, LD ;
Bassi, FH ;
Laird, M ;
Johnson, M ;
Newman, C .
OPHTHALMOLOGY, 1998, 105 (10) :1907-1914