Improved water and lipid suppression for 3D PRESS CSI using RF band selective inversion with gradient dephasing (BASING)

被引:161
作者
StarLack, J [1 ]
Nelson, SJ [1 ]
Kurhanewicz, J [1 ]
Huang, LR [1 ]
Vigneron, DB [1 ]
机构
[1] UNIV CALIF SAN FRANCISCO, DEPT RADIOL, MAGNET RESONANCE SCI CTR, SAN FRANCISCO, CA 94143 USA
关键词
magnetic resonance spectroscopy (MRS); water and lipid suppression; H-1 PRESS CSI; RF pulse design;
D O I
10.1002/mrm.1910380222
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
A T-1 insensitive solvent suppression technique-band selective inversion with gradient dephasing (BASING)-was developed to suppress water and lipids for H-1 magnetic resonance spectroscopy (MRS), BASING, which consists of a frequency selective RF inversion pulse surrounded by spoiler gradient pulses of opposite signs, was used to dephase stopband resonances and minimally impact passband metabolites, Passband phase linearity was achieved with a dual BASING scheme, Using the Shinnar-Le Roux algorithm, a highpass filter was designed to suppress water and rephase the lactate methyl doublet independently of TE, and water/lipid bandstop filters were designed for the brain and prostate, Phantom and in vivo experimental 3D PRESS CSI data were acquired at 1.5 T to compare BASING with CHESS and STIR suppression, With BASING, the measured suppression factor was over 100 times higher than with CHESS or STIR causing baseline distortions to be removed, It was shown that BASING can be incorporated into a variety of sequences to offer improved suppression in the presence of B-1 and T-1 inhomogeneites.
引用
收藏
页码:311 / 321
页数:11
相关论文
共 41 条
  • [1] METABOLISM OF HUMAN GLIOMAS - ASSESSMENT WITH H-1 MR SPECTROSCOPY AND F-18 FLUORODEOXYGLUCOSE PET
    ALGER, JR
    FRANK, JA
    BIZZI, A
    FULHAM, MJ
    DESOUZA, BX
    DUHANEY, MO
    INSCOE, SW
    BLACK, JL
    VANZIJL, PCM
    MOONEN, CTW
    DICHIRO, G
    [J]. RADIOLOGY, 1990, 177 (03) : 633 - 641
  • [2] SPATIAL LOCALIZATION IN NMR-SPECTROSCOPY INVIVO
    BOTTOMLEY, PA
    [J]. ANNALS OF THE NEW YORK ACADEMY OF SCIENCES, 1987, 508 : 333 - 348
  • [3] NMR CHEMICAL-SHIFT IMAGING IN 3 DIMENSIONS
    BROWN, TR
    KINCAID, BM
    UGURBIL, K
    [J]. PROCEEDINGS OF THE NATIONAL ACADEMY OF SCIENCES OF THE UNITED STATES OF AMERICA-BIOLOGICAL SCIENCES, 1982, 79 (11): : 3523 - 3526
  • [4] NONINVASIVE DIFFERENTIATION OF TUMORS WITH USE OF LOCALIZED H-1 MR SPECTROSCOPY INVIVO - INITIAL EXPERIENCE IN PATIENTS WITH CEREBRAL-TUMORS
    BRUHN, H
    FRAHM, J
    GYNGELL, ML
    MERBOLDT, KD
    HANICKE, W
    SAUTER, R
    HAMBURGER, C
    [J]. RADIOLOGY, 1989, 172 (02) : 541 - 548
  • [5] MR IMAGING - CLINICAL USE OF THE INVERSION RECOVERY SEQUENCE
    BYDDER, GM
    YOUNG, IR
    [J]. JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 1985, 9 (04) : 659 - 675
  • [6] LOCALIZED IN-VIVO H-1 MAGNETIC-RESONANCE SPECTROSCOPY AND IN-VITRO ANALYSES OF HETEROGENEOUS BRAIN-TUMORS
    CHANG, L
    BOOTH, RA
    MCBRIDE, D
    BUCHTHAL, SD
    MILLER, BL
    ERNST, T
    CORNFORD, M
    JENDEN, D
    [J]. JOURNAL OF NEUROIMAGING, 1995, 5 (03) : 157 - 163
  • [7] Single and multiple-selective excitation combined with pulsed field gradients
    Dalvit, C
    Ko, SY
    Bohlen, JM
    [J]. JOURNAL OF MAGNETIC RESONANCE SERIES B, 1996, 110 (02): : 124 - 131
  • [8] 3D PHASE ENCODING H-1 SPECTROSCOPIC IMAGING OF HUMAN BRAIN
    DUIJN, JH
    MATSON, GB
    MAUDSLEY, AA
    WEINER, MW
    [J]. MAGNETIC RESONANCE IMAGING, 1992, 10 (02) : 315 - 319
  • [9] IMPROVED WATER SUPPRESSION FOR LOCALIZED IN-VIVO H-1 SPECTROSCOPY
    ERNST, T
    HENNIG, J
    [J]. JOURNAL OF MAGNETIC RESONANCE SERIES B, 1995, 106 (02): : 181 - 186
  • [10] LOCALIZED HIGH-RESOLUTION PROTON NMR-SPECTROSCOPY USING STIMULATED ECHOES - INITIAL APPLICATIONS TO HUMAN-BRAIN INVIVO
    FRAHM, J
    BRUHN, H
    GYNGELL, ML
    MERBOLDT, KD
    HANICKE, W
    SAUTER, R
    [J]. MAGNETIC RESONANCE IN MEDICINE, 1989, 9 (01) : 79 - 93