A comparison of hemiarthroplasty with a novel polycarbonate-urethane acetabular component for displaced intracapsular fractures of the femoral neck A RANDOMISED CONTROLLED TRIAL IN ELDERLY PATIENTS

被引:55
作者
Cadossi, M. [1 ]
Chiarello, E. [1 ]
Savarino, L. [1 ]
Tedesco, G. [1 ]
Baldini, N. [1 ]
Faldini, C. [1 ]
Giannini, S. [1 ]
机构
[1] Rizzoli Orthopaed Inst, Bologna, Italy
关键词
TOTAL HIP-REPLACEMENT; ARTHROPLASTY; BEARING; MULTICENTER; PROSTHESIS; FIXATION;
D O I
10.1302/0301-620X.95B5.31083
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
100224 [整形外科学];
摘要
We undertook a randomised controlled trial to compare bipolar hemiarthroplasty (HA) with a novel total hip replacement (THR) comprising a polycarbonate-urethane (PCU) acetabular component coupled with a large-diameter metal femoral head for the treatment of displaced fractures of the femoral neck in elderly patients. Functional outcome, assessed with the Harris hip score (HHS) at three months and then annually after surgery, was the primary endpoint. Rates of revision and complication were secondary endpoints. Based on a power analysis, 96 consecutive patients aged >70 years were randomised to receive either HA (49) or a PCU-THR (47). The mean follow-up was 30.1 months (23 to 50) and 28.6 months (22 to 52) for the HA and the PCU group, respectively. The HHS showed no statistically significant difference between the groups at every follow-up. Higher pain was recorded in the PCU group at one and two years' follow-up (p = 0.006 and p = 0.019, respectively). In the HA group no revision was performed. In the PCU-THR group six patients underwent revision and one patient is currently awaiting re-operation. The three-year survival rate of the PCU-THR group was 0.841 (95% confidence interval 0.680 to 0.926). Based on our findings we do not recommend the use of the PCU acetabular component as part of the treatment of patients with fractures of the femoral neck.
引用
收藏
页码:609 / 615
页数:7
相关论文
共 22 条
[1]
Total hip arthroplasty and hemiarthroplasty in mobile, independent patients with a displaced intracapsular fracture of the femoral neck - A randomized, controlled trial [J].
Baker, R. P. ;
Squires, B. ;
Gargan, M. F. ;
Bannister, G. C. .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2006, 88A (12) :2583-2589
[2]
Operative management of displaced femoral neck fractures in elderly patients - An international survey [J].
Bhandari, M ;
Devereaux, PJ ;
Tornetta, P ;
Swiontkowski, MF ;
Berry, DJ ;
Haidukewych, G ;
Schemitsch, EH ;
Hanson, BP ;
Koval, K ;
Dirschl, D ;
Leece, P ;
Keel, M ;
Petrisor, B ;
Heetveld, M ;
Guyatt, GH .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2005, 87A (09) :2122-2130
[3]
Internal fixation compared with arthroplasty for displaced fractures of the femoral neck - A meta-analysis [J].
Bhandari, M ;
Devereaux, PJ ;
Swiontkowski, MF ;
Tornetta, P ;
Obremskey, W ;
Koval, KJ ;
Nork, S ;
Sprague, S ;
Schemitsch, EH ;
Guyatt, GH .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2003, 85A (09) :1673-1681
[4]
A randomised controlled trial comparing bipolar hemiarthroplasty with total hip replacement for displaced intracapsular fractures of the femoral neck in elderly patients [J].
Blomfeldt, R. ;
Tornkvist, H. ;
Eriksson, K. ;
Soderqvist, A. ;
Ponzer, S. ;
Tidermark, J. .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 2007, 89B (02) :160-165
[5]
ECTOPIC OSSIFICATION FOLLOWING TOTAL HIP-REPLACEMENT - INCIDENCE AND A METHOD OF CLASSIFICATION [J].
BROOKER, AF ;
BOWERMAN, JW ;
ROBINSON, RA ;
RILEY, LH .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1973, A 55 (08) :1629-1632
[6]
Aging performance of a compliant layer bearing acetabular prosthesis in an ovine hip arthroplasty model [J].
Carbone, Angelo ;
Howie, Donald W. ;
McGee, Margaret ;
Field, John ;
Pearcy, Mark ;
Smith, Nigel ;
Jones, Eric .
JOURNAL OF ARTHROPLASTY, 2006, 21 (06) :899-906
[7]
Long-Term Evaluation of a Compliant Cushion Form Acetabular Bearing for Hip Joint Replacement: A 20 Million Cycles Wear Simulation [J].
Elsner, Jonathan J. ;
Shemesh, Maoz ;
Mezape, Yoav ;
Levenshtein, Mario ;
Hakshur, Keren ;
Shterling, Avi ;
Linder-Ganz, Eran ;
Eliaz, Noam .
JOURNAL OF ORTHOPAEDIC RESEARCH, 2011, 29 (12) :1859-1866
[8]
Wear rate evaluation of a novel polycarbonate-urethane cushion form bearing for artificial hip joints [J].
Elsner, Jonathan J. ;
Mezape, Yoav ;
Hakshur, Keren ;
Shemesh, Maoz ;
Linder-Ganz, Eran ;
Shterling, Avi ;
Eliaz, Noam .
ACTA BIOMATERIALIA, 2010, 6 (12) :4698-4707
[9]
LOW-ANGLE FIXATION IN FRACTURES OF THE FEMORAL NECK [J].
GARDEN, RS .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1961, 43 (04) :647-663
[10]
Giannini S, 2011, AGING CLIN EXP RES, V23, P40