The influence of finite aperture and frequency response of ultrasonic hydrophone probes on the determination of acoustic output

被引:8
作者
Radulescu, EG
Lewin, PA
Wójcik, J
Nowicki, A
Berger, WA
机构
[1] Drexel Univ, Sch Biomed Engn, Dept Elect & Comp Engn, Philadelphia, PA 19104 USA
[2] Polish Acad Sci, IPPT, PL-00049 Warsaw, Poland
[3] Univ Scranton, Scranton, PA 18510 USA
关键词
ultrasound imaging; nonlinear propagation; spatial averaging; safety indices;
D O I
10.1016/j.ultras.2003.11.019
中图分类号
O42 [声学];
学科分类号
070206 ; 082403 ;
摘要
The influence of finite aperture and frequency response of piezoelectric ultrasonic hydrophone probes on the Thermal and Mechanical Indices was investigated using a comprehensive acoustic wave propagation model. The experimental verification of the model was obtained using a commercially available, 8 MHz, dynamically focused linear array and a single element, 5 MHz, focused rectangular source. The pressure-time waveforms were recorded using piezoelectric polymer hydrophone probes of different active element diameters and bandwidths. The nominal diameters of the probes ranged from 50 to 500 mum and their usable bandwidths varied between 55 and 100 MHz. The Pulse Intensity Integral (PII), used to calculate the Thermal Index (TI), was found to increase with increasing bandwidth and decreasing effective aperture of the probes. The Mechanical Index (MI), another safety indicator, was also affected, but to a lesser extent. The corrections needed were predicted using the model and successfully reduced the discrepancy as large as 30% in the determination of PII. The results of this work indicate that by accounting for hydrophones' finite aperture and correcting the value of PII, all intensities derived from the PII can be corrected for spatial averaging error. The results also point out that a caution should be exercised when comparing acoustic output data. In particular, hydrophone's frequency characteristics of the effective diameter and sensitivity are needed to correctly determine the MI, TI, and the total acoustic output power produced by an imaging transducer. (C) 2003 Elsevier B.V. All rights reserved.
引用
收藏
页码:367 / 372
页数:6
相关论文
共 9 条
[1]  
*AIUM, 1998, AC OUTP MEAS STAND D
[2]  
American Institute of Ultrasound in Medicine/ National Electrical Manufacturers Association (AIUM/ NEMA), 1998, STAND REAL TIM DISPL
[3]  
Filipczynski L., 2001, Archives of Acoustics, V26, P223
[4]   Hydrophones' effective diameter measurements as a quasi-continuous function of frequency [J].
Radulescu, EG ;
Lewin, PA ;
Nowicki, A ;
Berger, WA .
ULTRASONICS, 2003, 41 (08) :635-641
[5]   Calibration of ultrasonic hydrophone probes up to 100 MHz using time gating frequency analysis and finite amplitude waves [J].
Radulescu, EG ;
Lewin, PA ;
Wójcik, J ;
Nowicki, A .
ULTRASONICS, 2003, 41 (04) :247-254
[6]   Nonlinear propagation model for ultrasound hydrophones calibration in the frequency range up to 100 MHz [J].
Radulescu, EG ;
Wójcik, J ;
Lewin, PA ;
Nowicki, A .
ULTRASONICS, 2003, 41 (04) :239-245
[7]   1-60 MHz measurements in focused acoustic fields using spatial averaging corrections [J].
Radulescu, EG ;
Lewin, PA ;
Nowicki, A .
ULTRASONICS, 2002, 40 (1-8) :497-501
[8]  
RADULESCU EG, 2001, IEEE T UFFC, V46, P1575
[9]   Conservation of energy and absorption in acoustic fields for linear and nonlinear propagation [J].
Wójcik, J .