Experimental study on heat production by a 23.5-kHz ultrasonically activated device for endoscopic surgery

被引:59
作者
Kinoshita, T [1 ]
Kanehira, E [1 ]
Omura, K [1 ]
Kawakami, K [1 ]
Watanabe, Y [1 ]
机构
[1] Kanazawa Univ, Sch Med, Dept Surg 1, Kanazawa, Ishikawa 9208641, Japan
来源
SURGICAL ENDOSCOPY-ULTRASOUND AND INTERVENTIONAL TECHNIQUES | 1999年 / 13卷 / 06期
关键词
ultrasonically activated device; thermal damage; endoscopic surgery;
D O I
10.1007/s004649901055
中图分类号
R61 [外科手术学];
学科分类号
摘要
An experimental study was carried out to evaluate heat production by an ultrasonically activated device (USAD) using an animal model. In an anesthetized living pig, the gastroepiploic and mesenteric vessels were coagulated and cut by an USAD at a power level of 70% (n = 8) or 100% (n = 8). During the division, the time-discrete temperature change on the surface of the animal tissue adjacent to the blade was measured by thermography. To compare the USAD with conventional electrocautery (EC), a full-thickness incision of the gastric wall was performed by each device, and the temperature change was measured. With the USAD, the temperature increased gradually and remained below 150 degrees C during the entire activating time at both power levels. By contrast, with EC at 30 W, the temperature increased rapidly and exceeded 350 degrees C within only a few seconds. The area above 60 degrees C reached a final width of 10 mm for the USAD, as compared with 22 mm for EC. Microscopically, thermal alterations such as carbonization and vaporization were much more severe and extensive in the adjacent tissue when using EC rather than the USAD. With the USAD, heat production is much slower and more limited than with conventional EC; thus, the USAD causes fewer thermal alterations in adjacent tissue. USAD should be preferred for tissue coagulation and cutting during endscopic surgery.
引用
收藏
页码:621 / 625
页数:5
相关论文
共 10 条
[1]   Experimental comparison of the ultrasonically-activated scalpel to electrosurgery and laser surgery for laparoscopic use [J].
Amaral, JF ;
Chrostek, CA .
MINIMALLY INVASIVE THERAPY & ALLIED TECHNOLOGIES, 1997, 6 (04) :324-331
[2]  
AMARAL JF, 1995, SURG LAPAROSC ENDOSC, V5, P255
[3]  
AMARAL JF, 1994, SURG LAPAROSC ENDOSC, V4, P92
[4]  
HAMBLEY R, 1988, J DERMATOL SURG ONC, V14, P11
[5]   Proximity injury by the ultrasonically activated scalpel during dissection [J].
Kadesky, KM ;
Schopf, B ;
Magee, JF ;
Blair, GK .
JOURNAL OF PEDIATRIC SURGERY, 1997, 32 (06) :878-879
[6]   Development of a 23.5 kHz ultrasonically activated device for laparoscopic surgery [J].
Kanehira, E ;
Omura, K ;
Kinoshita, T ;
Sasaki, M ;
Watanabe, T ;
Kawakami, K ;
Watanabe, Y .
MINIMALLY INVASIVE THERAPY & ALLIED TECHNOLOGIES, 1998, 7 (04) :315-319
[7]   New technology for the division of short gastric vessels during laparoscopic Nissen fundoplication - A prospective randomized trial [J].
Laycock, WS ;
Trus, TL ;
Hunter, JG .
SURGICAL ENDOSCOPY-ULTRASOUND AND INTERVENTIONAL TECHNIQUES, 1996, 10 (01) :71-73
[8]  
Mueller W, 1994, Endosc Surg Allied Technol, V2, P205
[9]  
REIDENBACH HD, 1992, OPERATIVE MANUAL END, P46
[10]   The use of bipolar cautery, laparosonic coagulating shears, and vascular clips for hemostasis of small and medium-sized vessels [J].
Spivak, H ;
Richardson, WS ;
Hunter, JG .
SURGICAL ENDOSCOPY-ULTRASOUND AND INTERVENTIONAL TECHNIQUES, 1998, 12 (02) :183-185