Construction and characterization of a high-flow, high-resolution ion mobility spectrometer for detection of explosives after personnel portal sampling

被引:43
作者
Wu, C
Steiner, WE
Tornatore, PS
Matz, LM
Siems, WF
Atkinson, DA
Hill, HH [1 ]
机构
[1] Washington State Univ, Dept Chem, Pullman, WA 99164 USA
[2] Ion Track Instruments, Wilmington, MA 01887 USA
[3] Idaho Natl Lab, Dept Chem, Idaho Falls, ID 83401 USA
关键词
ion mobility spectrometry; high-flow; high-resolution; personnel portal;
D O I
10.1016/S0039-9140(01)00680-4
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A novel analysis of explosives via the coupling of an airline passenger personnel portal with a high-flow (HF), high-resolution (HR) ion mobility spectrometry (IMS) was shown for the first time. The HF-HR-IMS utilized a novel ion aperture grid design with a Ni-63 ionization source while operating at ambient pressure in the positive ion mode at 200 degreesC. The HF-HR-IMS response characteristics of 2,4,6-trinitrotoluene (TNT), 4,6-dinitro-o-cresol (4,6DNOC), and cyclo-1,3,5-trimethylene-2,4,6-trinitramine (RDX) were investigated. Modifications made to the HF-HR-IMS exhaust and ionization source created an 800% increase in the total ion current (TIC), from 0.85 to 6.8 nA. This translated into a 65% ion response increase for TNT when compared with a traditional IMS. A mixture of TNT and (4,6DNOC) was used to successfully demonstrate the resolving power of the species with similar reduced mobility constants (K-o), 1.54 and 1.59, respectively. The reactant ion (H2O)(n)H+, peak was also used to measure the resolving power of the spectrometer while varying the internal diameter of three different aperture openings from 1.00 to 3.54cm. This provided a resolving power range of 50-60, double that typically achievable by commercial IMS instruments. Most important, these changes made in this new instrumental design can be implemented to all existing and future IMS's to greatly enhance the achievable IMS resolving power. (C) 2002 Elsevier Science B.V. All rights reserved.
引用
收藏
页码:123 / 134
页数:12
相关论文
共 21 条
[1]   Characterization of tear gas residues by ion mobility spectrometry [J].
Allison, G ;
McLeod, CW .
APPLIED SPECTROSCOPY, 1997, 51 (12) :1880-1889
[2]   TUNABLE SELECTIVE DETECTION FOR CAPILLARY GAS-CHROMATOGRAPHY BY ION MOBILITY MONITORING [J].
BAIM, MA ;
HILL, HH .
ANALYTICAL CHEMISTRY, 1982, 54 (01) :38-43
[3]  
CLEMENT RE, 1992, INSTRUMENTATION TRAC
[4]   PLASMA CHROMATOGRAPHY - A NEW DIMENSION FOR GAS CHROMATOGRAPHY AND MASS SPECTROMETRY [J].
COHEN, MJ ;
KARASEK, FW .
JOURNAL OF CHROMATOGRAPHIC SCIENCE, 1970, 8 (06) :330-&
[5]  
EICEMAN GA, 1989, ANAL INSTRUM, V18, P227
[6]  
FYTCHE LM, 1992, J FORENSIC SCI, V37, P1550
[7]  
Garofolo F, 1996, RAPID COMMUN MASS SP, V10, P1321, DOI 10.1002/(SICI)1097-0231(199608)10:11<1321::AID-RCM556>3.3.CO
[8]  
2-W
[9]  
GERMAN J, 1997, SANDIA LAB NEWS 0916
[10]   PLASMA CHROMATOGRAPHY [J].
KARASEK, FW .
ANALYTICAL CHEMISTRY, 1974, 46 (08) :A710-+