Computerized localization of breast lesions from two views - An experimental comparison of two methods

被引:18
作者
Chang, YH [1 ]
Good, WF [1 ]
Sumkin, JH [1 ]
Zheng, B [1 ]
Gur, D [1 ]
机构
[1] Univ Pittsburgh, Dept Radiol, Pittsburgh, PA 15261 USA
关键词
computer-aided diagnosis; localization (triangulation); receiver operating characteristic methodology;
D O I
10.1097/00004424-199909000-00006
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
RATIONALE AND OBJECTIVES. The authors compared two computerized methods, the are and cartesian straight- line, for the localization of breast lesions in two mammographic views. METHODS. A total Of 571 craniocaudal and 571 mediolateral oblique matched mammographic image pairs (or 1142 individual images) depicting 290 pathology-verified masses on both views were selected from our image database. Using a previously developed computer-aided detection scheme, all 290 masses and 3992 suspicious but negative regions were identified. After pairing all identified regions from both views, all masses (true-positive-true-positive matched pairs) and a total of 10330 false-positive pairs (including false-positive-false-positive, true-positive-false-positive, and false-positive-true positive pairs) were assessed as to their position in relation to the nipple using both the are and the cartesian straight-line methods. Receiver operating characteristic methodology was used to evaluate the performance levels for each method in determining, based solely on location, whether a pair of suspicious regions represented a true mass or a false-positive combination, RESULTS. The areas under the receiver operating characteristic curves (A(z)) were 0.79 and 0.78 for the arc and cartesian straight-line methods, respectively. The difference between the two techniques las measured by A(z)) was not statistically significant (P > 0.99). CONCLUSIONS. These preliminary results demonstrated that the two methods are comparable in identifying true masses from triangulated observations on two views, However, the are method is somewhat favorable because only the nipple location is required for localization.
引用
收藏
页码:585 / 588
页数:4
相关论文
共 13 条
[1]   A simple method for automatically locating the nipple on mammograms [J].
Chandrasekhar, R ;
Attikiouzel, Y .
IEEE TRANSACTIONS ON MEDICAL IMAGING, 1997, 16 (05) :483-494
[2]  
FOLIO LR, 1994, APPL RADIOL DEC, P17
[3]   Multi-Image CAD employing features derived from ipsilateral mammographic views [J].
Good, WF ;
Zheng, B ;
Chang, YH ;
Wang, XH ;
Maitz, G ;
Gur, D .
MEDICAL IMAGING 1999: IMAGE PROCESSING, PTS 1 AND 2, 1999, 3661 :474-485
[4]  
GOOD WF, 1999, IN PRESS P 13 INT C
[5]  
Kopans D. B., 1997, BREAST IMAGING
[6]   LOCALIZATION OF BREAST-LESIONS IDENTIFIED ON ONLY ONE MAMMOGRAPHIC VIEW [J].
KOPANS, DB ;
WAITZKIN, ED ;
LINETSKY, L ;
SWANN, CA ;
MCCARTHY, KA ;
HALL, DA ;
WHITE, G .
AMERICAN JOURNAL OF ROENTGENOLOGY, 1987, 149 (01) :39-41
[8]   Incorporation of a set enumeration trees-based classifier into a hybrid computer-assisted diagnosis scheme for mass detection [J].
Rymon, R ;
Zheng, B ;
Chang, YH ;
Gur, D .
ACADEMIC RADIOLOGY, 1998, 5 (03) :181-187
[9]   LOCALIZATION OF OCCULT BREAST-LESIONS - PRACTICAL SOLUTIONS TO PROBLEMS OF TRIANGULATION [J].
SWANN, CA ;
KOPANS, DB ;
MCCARTHY, KA ;
WHITE, G ;
HALL, DA .
RADIOLOGY, 1987, 163 (02) :577-579
[10]   Adequacy testing of training set sample sizes in the development of a computer-assisted diagnosis scheme [J].
Zheng, B ;
Chang, YH ;
Good, WF ;
Gur, D .
ACADEMIC RADIOLOGY, 1997, 4 (07) :497-502