A preliminary laboratory investigation of air embolus detection and grading using an artificial neural network

被引:1
作者
Strong, K [1 ]
Westenskow, DR [1 ]
Fine, PG [1 ]
Orr, JA [1 ]
机构
[1] UNIV UTAH,HLTH SCI CTR,DEPT ANESTHESIOL 3C444,SALT LAKE CITY,UT 84132
来源
INTERNATIONAL JOURNAL OF CLINICAL MONITORING AND COMPUTING | 1997年 / 14卷 / 02期
关键词
laboratory study; neural network;
D O I
10.1023/A:1016925306538
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
Processed digitized Doppler signals abstracted from recordings during continuous air infusion in dogs were used to train a neural network to estimate air embolism infusion rates. Background: Precordial Doppler is a sensitive technique for detecting venous air embolism during anesthesia, but it requires constant attentive listening. Since neural networks are particularly well suited to the task of pattern recognition, we sought to investigate this technology for detection and grading of air embolism. Methods: Air was infused into peripheral veins of four anesthetized dogs at rates of 0.025, 0.05, 0.10, 0.25, 0.50 and 1.0 ml(-1).kg(-1).min(-1) while digital recordings of the precordial Doppler ultrasound signal were collected. The frequency content of the recordings was determined by Fourier analysis. The output of the Fourier transform was the input to a neural network. The network was then trained to estimate the air infusion rate. Results: The correlation coefficient between the size of the air embolism and the air infusion rate was greater than r(2) = 0.93 for each of the four animals in the study when the network was trained using the data for all four dogs. When the data from a dog was withheld from the training set and used only for testing the correlation coefficients ranged from r(2) = 0.75 to r(2) = 0.27. For frequencies below 250 Hz, the acoustic energy tended to fall as the air infusion rate increased. The opposite occurred at frequencies above 325 Hz. Conclusions: Neural network processing of the precordial Doppler signal provides a quantitative estimate of the size of an air embolism.
引用
收藏
页码:103 / 107
页数:5
相关论文
共 25 条
[1]  
ALTOBELLI SA, 1985, CARDIOVASCULAR ULTRA
[2]  
BARASH PG, 1989, CLIN ANESTH, P859
[3]  
Christensen D., 1988, ULTRASONIC BIOINSTRU
[4]   COMPARISON OF VENOUS AIR-EMBOLISM MONITORING METHODS IN SUPINE DOGS [J].
ENGLISH, JB ;
WESTENSKOW, D ;
HODGES, MR ;
STANLEY, TH .
ANESTHESIOLOGY, 1978, 48 (06) :425-429
[5]  
FINE PG, 1991, ANESTHESIOLOGY S, V75, pA465
[6]   AIR-EMBOLISM DURING POSTERIOR SPINAL-FUSION [J].
FRANKEL, AS ;
HOLZMAN, RS .
CANADIAN JOURNAL OF ANAESTHESIA-JOURNAL CANADIEN D ANESTHESIE, 1988, 35 (05) :511-514
[7]   COMPUTER-ASSISTED DOPPLER MONITORING TO ENHANCE DETECTION OF AIR EMBOLI [J].
GIBBY, GL ;
GHANI, GA .
JOURNAL OF CLINICAL MONITORING, 1988, 4 (01) :64-73
[8]   THE EFFICACY OF DOPPLER MONITORING FOR THE DETECTION OF VENOUS AIR-EMBOLISM [J].
GILDENBERG, PL ;
OBRIEN, RP ;
BRITT, WJ ;
FROST, EAM .
JOURNAL OF NEUROSURGERY, 1981, 54 (01) :75-78
[9]   INCIDENCE AND CARDIAC EFFECTS OF SYSTEMIC VENOUS AIR-EMBOLISM - ECHOCARDIOGRAPHIC EVIDENCE OF ARTERIAL EMBOLIZATION VIA NONCARDIAC SHUNT [J].
GOTTDIENER, JS ;
PAPADEMETRIOU, V ;
NOTARGIACOMO, A ;
PARK, WY ;
CUTLER, J .
ARCHIVES OF INTERNAL MEDICINE, 1988, 148 (04) :795-800
[10]  
GRAVENSTEIN JS, 1982, MONITORING PRACTICE, P279