Amperometric biosensor for ethanol based on immobilization of alcohol dehydrogenase on a nickel hexacyanoferrate modified microband gold electrode

被引:76
作者
Cai, CX
Xue, KH
Zhou, YM
Yang, H
机构
[1] Department of Chemistry, Nanjing Normal University
关键词
alcohol dehydrogenase; amperometry; biosensors; electrocatalytic oxidation; NADH;
D O I
10.1016/S0039-9140(96)02049-8
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Alcohol dehydrogenase (ADH) has been immobilized on a nickel hexacyanoferrate modified microband gold electrode surface by a glutaraldehyde/bovine serum albumin (BSA) cross-linking procedure to provide a new amperometric sensor for the assay of ethanol. The resulting enzyme electrode exhibits excellent electrocatalysis for the oxidation of reduced nicotinamide-adenine dinucleotide (NADH). The amperometric determination is based on the electrochemical detection of NADH which is generated in the enzymatic reaction of ethanol with NAD(+) under catalysis of ADH. The influence of various experimental conditions was examined for the determination of the optimum analytical performance. The sensor responds rapidly to ethanol with a detection limit of (5.0 +/- 0.3) x 10(-7) mol l(-1). The response current increases linearly with ethanol concentration up to 5 mmol l(-1). The sensor remains relatively stable for about 1 week. (C) 1997 Published by Elsevier Science B.V.
引用
收藏
页码:339 / 347
页数:9
相关论文
共 49 条
[1]   AN ORGANIC CONDUCTOR ELECTRODE FOR THE OXIDATION OF NADH [J].
ALBERY, WJ ;
BARTLETT, PN .
JOURNAL OF THE CHEMICAL SOCIETY-CHEMICAL COMMUNICATIONS, 1984, (04) :234-236
[2]   VOLTAMMETRIC STUDIES OF THE OXIDATION OF REDUCED NICOTINAMIDE ADENINE-DINUCLEOTIDE AT A CONDUCTING POLYMER ELECTRODE [J].
ATTA, NF ;
GALAL, A ;
KARAGOZLER, AE ;
ZIMMER, H ;
RUBINSON, JF ;
MARK, HB .
JOURNAL OF THE CHEMICAL SOCIETY-CHEMICAL COMMUNICATIONS, 1990, (19) :1347-1349
[3]  
Bernadette F. Y., 1987, ANAL CHEM, V59, P2111
[4]   CATALYTIC-OXIDATION OF REDUCED NICOTINAMIDE ADENINE-DINUCLEOTIDE AT A MICROBAND GOLD ELECTRODE MODIFIED WITH NICKEL HEXACYANOFERRATE [J].
CAI, CX ;
JU, HX ;
CHEN, HY .
ANALYTICA CHIMICA ACTA, 1995, 310 (01) :145-151
[5]  
CAI CX, 1995, ACTA CHIM SINICA, V53, P281
[6]   ELECTROCATALYTIC OXIDATION AND LIQUID-CHROMATOGRAPHIC DETECTION OF ALIPHATIC-ALCOHOLS AT A NICKEL-BASED GLASSY-CARBON MODIFIED ELECTRODE [J].
CASELLA, IG ;
CATALDI, TRI ;
SALVI, AM ;
DESIMONI, E .
ANALYTICAL CHEMISTRY, 1993, 65 (21) :3143-3150
[7]   MIXED-VALENT RUTHENIUM OXIDE RUTHENIUM CYANIDE INORGANIC FILM ON GLASSY-CARBON ELECTRODES AS AN AMPEROMETRIC SENSOR OF ALIPHATIC-ALCOHOLS [J].
CATALDI, TRI ;
CENTONZE, D ;
GUERRIERI, A .
ANALYTICAL CHEMISTRY, 1995, 67 (01) :101-107
[8]   ELECTROCATALYSIS AND AMPEROMETRIC DETECTION OF ETHANOL AT RUTHENIUM-BASED INORGANIC FILMS WITH IMPROVED RESPONSE STABILITY [J].
CATALDI, TRI ;
CENTONZE, D ;
DESIMONI, E ;
FORASTIERO, V .
ANALYTICA CHIMICA ACTA, 1995, 310 (02) :257-262
[9]   ELECTROCATALYTIC OXIDATION AND FLOW-INJECTION DETERMINATION OF REDUCED NICOTINAMIDE COENZYME AT A GLASSY-CARBON ELECTRODE MODIFIED BY A POLYMER THIN-FILM [J].
CHI, QJ ;
DONG, SJ .
ANALYST, 1994, 119 (05) :1063-1066
[10]   COMPARISON OF SPECTROPHOTOMETRIC AND AMPEROMETRIC RATE PARAMETERS OF ENZYMATIC-REACTIONS [J].